Measurement of the Intensity of Long-Lifetime Luminophores in the Presence of Background Signals Using Phase-Modulation Fluorometry

Overview
Journal Appl Spectrosc
Specialties Chemistry
Pathology
Date 2020 Apr 15
PMID 32284628
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

We describe a method to correct for background fluorescence and to measure the intensity of long-lifetime probes using phase-modulation fluorometry. The theoretically predicted and simulated data were supported by two experiments. The fractional contribution of background fluorescence in the sample was determined by measurement of phase angle and/or modulation at single modulation frequency. In certain experimental situations, where the mean decay times of the background and the long lifetime probe are widely separated, determination of background signals in the sample does not require a blank sample or information about the nature of intensity decay of the background. Hence, phase-modulation fluorometry can be used to directly determine the intensity of the long-lifetime probe in the presence of an unknown short-lifetime background. We also discuss the effects of ambient light (indefinitely long lifetime) and scattered excitation light (zero lifetime) on phase-modulation measurements.

Citing Articles

Self-Referenced Multifrequency Phase-Resolved Luminescence Spectroscopy.

de la Torre A, Medina-Rodriguez S, Segura J, Fernandez-Sanchez J Sensors (Basel). 2020; 20(19).

PMID: 32987919 PMC: 7583794. DOI: 10.3390/s20195482.


Plasmon-controlled fluorescence towards high-sensitivity optical sensing.

Ray K, Chowdhury M, Zhang J, Fu Y, Szmacinski H, Nowaczyk K Adv Biochem Eng Biotechnol. 2008; 116:29-72.

PMID: 19082931 PMC: 6781866. DOI: 10.1007/10_2008_9.


A novel method for monitoring monoclonal antibody production during cell culture.

Szmacinski H, Smith D, Hanson M, Kostov Y, Lakowicz J, Rao G Biotechnol Bioeng. 2008; 100(3):448-57.

PMID: 18438872 PMC: 2729163. DOI: 10.1002/bit.21795.

References
1.
Kar S, Arnold M . Fiber-optic ammonia sensor for measuring synaptic glutamate and extracellular ammonia. Anal Chem. 1992; 64(20):2438-43. DOI: 10.1021/ac00044a024. View

2.
Murtaza Z, Chang Q, Rao G, Lin H, Lakowicz J . Long-lifetime metal-ligand pH probe. Anal Biochem. 1997; 247(2):216-22. PMC: 6816250. DOI: 10.1006/abio.1997.2057. View

3.
Hemmila I, Dakubu S, Mukkala V, Siitari H, Lovgren T . Europium as a label in time-resolved immunofluorometric assays. Anal Biochem. 1984; 137(2):335-43. DOI: 10.1016/0003-2697(84)90095-2. View

4.
Reinhart G, Marzola P, Jameson D, Gratton E . A method for on-line background subtraction in frequency domain fluorometry. J Fluoresc. 2013; 1(3):153-62. DOI: 10.1007/BF00865362. View

5.
Anderson D, Guo B, Xu Y, Ng L, Kricka L, Skogerboe K . Clinical chemistry. Anal Chem. 1997; 69(12):165R-229R. DOI: 10.1021/a1970008p. View