Objective: The development of new endovascular technologies and techniques for mechanical thrombectomy in stroke has greatly relied on benchtop simulators. This paper presents an affordable, versatile, and realistic benchtop simulation model for stroke.

Methods: A test bed for embolic occlusion of cerebrovascular arteries and mechanical thrombectomy was developed with 3D-printed and commercially available cerebrovascular phantoms, a customized hydraulic system to generate physiological flow rate and pressure, and 2 types of embolus analogs (elastic and fragment-prone) capable of causing embolic occlusions under physiological flow.

Results: The test bed was highly versatile and allowed realistic, radiation-free mechanical thrombectomy for stroke due to large-vessel occlusion with rapid exchange of geometries and phantom types. Of the transparent cerebrovascular phantoms tested, the 3D-printed phantom was the easiest to manufacture, the glass model offered the best visibility of the interaction between embolus and thrombectomy device, and the flexible model most accurately mimicked the endovascular system during device navigation. None of the phantoms modeled branches smaller than 1 mm or perforating arteries, and none underwent realistic deformation or luminal collapse from device manipulation or vacuum. The hydraulic system created physiological flow rate and pressure leading to iatrogenic embolization during thrombectomy in all phantoms. Embolus analogs with known fabrication technique, structure, and tensile strength were introduced and consistently occluded the middle cerebral artery bifurcation under physiological flow, and their interaction with the device was accurately visualized.

Conclusions: The test bed presented in this study is a low-cost, comprehensive, realistic, and versatile platform that enabled high-quality analysis of embolus-device interaction in multiple cerebrovascular phantoms and embolus analogs.

Citing Articles

Safety Window for Effective Lesion Crossing in Patients With Chronic Thromboembolic Pulmonary Hypertension.

Perkins S, Funes M, Cheah D, Argenti C, Vinales J, Gordon D J Soc Cardiovasc Angiogr Interv. 2024; 3(8):102142.

PMID: 39166161 PMC: 11330921. DOI: 10.1016/j.jscai.2024.102142.


Evaluating the accuracy of cerebrovascular computational fluid dynamics modeling through time-resolved experimental validation.

Luisi C, Witter T, Nikoubashman O, Wiesmann M, Steinseifer U, Neidlin M Sci Rep. 2024; 14(1):8194.

PMID: 38589554 PMC: 11001858. DOI: 10.1038/s41598-024-58925-8.


Clinical situations for which 3D printing is considered an appropriate representation or extension of data contained in a medical imaging examination: neurosurgical and otolaryngologic conditions.

Ali A, Morris J, Decker S, Huang Y, Wake N, Rybicki F 3D Print Med. 2023; 9(1):33.

PMID: 38008795 PMC: 10680204. DOI: 10.1186/s41205-023-00192-w.


Feasibility of a silicone vascular phantom replicating real arterial contrast filling dynamics on cerebral angiography: An in-vitro pilot study.

Lee K, Sheen J, Yoon J, Kim M, Maeng J, Hwang S PLoS One. 2023; 18(1):e0280395.

PMID: 36649351 PMC: 9844885. DOI: 10.1371/journal.pone.0280395.


Construction and Validation of a Benchtop Model for Testing of Mechanical Thrombectomy Devices for Pulmonary Embolism.

Madhani S, Larco J, Liu Y, Abbasi M, Shahid A, Yasin O Cardiovasc Intervent Radiol. 2022; 46(3):385-391.

PMID: 36482095 DOI: 10.1007/s00270-022-03326-5.