Evaluation of a Metal Artifact Reduction Algorithm and an Optimization Filter in the Estimation of Peri-implant Dehiscence Defects by Using Cone Beam Computed Tomography: an In-vitro Study

Overview
Publisher Elsevier
Date 2020 Mar 22
PMID 32197878
Citations 9
Authors
Affiliations
Soon will be listed here.
Abstract

Objectives: The aim of this study was to assess the effect of a metal artifact reduction (MAR) algorithm and the adaptive image noise optimizer (AINO) optimization filter in the detection of peri-implant dehiscences with cone beam computed tomography (CBCT).

Study Design: Nine implants (3 zirconium, 3 titanium, and 3 zirconium-titanium) were placed in 3 sheep heads. Dehiscences were created on the buccal and lingual/palatal surfaces. A total of 9 defects and 9 controls with no defects were evaluated by 3 observers. Each sheep head was scanned 5 times with 4 scan modes; (1) without MAR/without AINO; (2) with MAR/without AINO; (3) without MAR/with AINO; and (4) with MAR/with AINO. Receiver operating characteristic analysis and weighted kappa coefficients were used to calculate diagnostic efficacy and intra- and interobserver agreements for each implant type and scan mode.

Results: For all implant types, dehiscences were most accurately detected when both MAR and AINO were applied (P ≤ .045). Detection of dehiscences was more accurate with titanium implants (P ≤ .040). There were no significant differences in agreement among and between the observers.

Conclusions: The use of both MAR and AINO enhanced the detection accuracy of artificially created dehiscences in proximity to implants. Their combined use is recommended for detecting peri-implant dehiscences.

Citing Articles

Assessment of periodontal dehiscence and fenestration using ultrasonography and cone-beam computed tomography: an in vitro study.

Seker C, Yildiran B, Haylaz E, Geduk G, Icen M Clin Oral Investig. 2024; 28(12):665.

PMID: 39604758 DOI: 10.1007/s00784-024-06011-8.


Performance of different cone-beam computed tomography scan modes with and without metal artifact reduction in detection of recurrent dental caries under various restorative materials.

Abesi F, Talachi F, Ezoji F Pol J Radiol. 2024; 89:e281-e291.

PMID: 39040560 PMC: 11262014. DOI: 10.5114/pjr/188257.


Diagnostic accuracy of cone-beam computed tomography with modified grayscale range for detection of buccal cortical plate defects adjacent to dental implants.

Yaghoobi S, Dalili Kajan Z, Khosravifard N, Khalighi Sigaroudi A, Jouybari R Dent Res J (Isfahan). 2023; 20:90.

PMID: 37810449 PMC: 10557990.


The Effect of Changes in the Angular Position of Implants on Metal Artifact Reduction in Cone-Beam Computed Tomography Images: A Scoping Review.

Asnaashari M, Sadeghipour M, Bahrani Z, Valizadeh S, Moshfeghi M Radiol Res Pract. 2023; 2023:5539719.

PMID: 37554657 PMC: 10406552. DOI: 10.1155/2023/5539719.


Efficacy of low-dose cone beam computed tomography and metal artifact reduction tool for assessment of peri-implant bone defects: an in vitro study.

Nomier A, Gaweesh Y, Taalab M, El Sadat S BMC Oral Health. 2022; 22(1):615.

PMID: 36528573 PMC: 9759909. DOI: 10.1186/s12903-022-02663-8.