Comparison of Implant Stability Measurements Between a Resonance Frequency Analysis Device and a Modified Damping Capacity Analysis Device: an Study

Overview
Specialty Dentistry
Date 2020 Mar 5
PMID 32128274
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: A stability-measuring device that utilizes damping capacity analysis (DCA) has recently been introduced in the field of dental implantology. This study aimed to evaluate the sensitivity and reliability of this device by measuring the implant stability of samples in comparison with a resonance frequency analysis (RFA) device.

Methods: Six implant beds were prepared in porcine ribs using 3 different drilling protocols to simulate various implant stability conditions. Thirty-six pork ribs and 216 bone-level implants measuring 10 mm in height were used. The implant beds were prepared using 1 of the following 3 drilling protocols: 10-mm drilling depth with a 3.5-mm-diameter twist drill, 5-mm drilling depth with a 4.0-mm-diameter twist drill, and 10-mm drilling depth with a 4.0-mm-diameter twist drill. The first 108 implants were external-connection implants 4.0 mm in diameter, while the other 108 implants were internal-connection implants 4.3 mm in diameter. The peak insertion torque (PIT) during implant placement, the stability values obtained with DCA and RFA devices after implant placement, and the peak removal torque (PRT) during implant removal were measured.

Results: The intraclass correlation coefficients (ICCs) of the implant stability quotient (ISQ) results obtained using the RFA device at the medial, distal, ventral, and dorsal points were 0.997, 0.994, 0.994, and 0.998, respectively. The ICCs of the implant stability test (IST) results obtained using the DCA device at the corresponding locations were 0.972, 0.975, 0.974, and 0.976, respectively. Logarithmic relationships between PIT and IST, PIT and ISQ, PRT and IST, and PRT and ISQ were observed. The mean absolute difference between the ISQ and IST values on a Bland-Altman plot was -6.76 (-25.05 to 11.53, <0.05).

Conclusions: Within the limits of studies, measurements made using the RFA and DCA devices were found to be correlated under a variety of stability conditions.

Citing Articles

Correlation of two different devices for the evaluation of primary implant stability depending on dental implant length and bone density: An in vitro study.

Lee J, Lim Y, Ahn J, Kim B, Baek Y, Lim B PLoS One. 2024; 19(5):e0290595.

PMID: 38776308 PMC: 11111012. DOI: 10.1371/journal.pone.0290595.


Advancements in Methods of Classification and Measurement Used to Assess Tooth Mobility: A Narrative Review.

Kim G, Kim S, Chang J, Pyo S J Clin Med. 2024; 13(1).

PMID: 38202149 PMC: 10779763. DOI: 10.3390/jcm13010142.


Tissue-engineered bone construct promotes early osseointegration of implants with low primary stability in oversized osteotomy.

Xu L, Jacobs R, Cao Y, Sun X, Qin X BMC Oral Health. 2024; 24(1):69.

PMID: 38200461 PMC: 10782778. DOI: 10.1186/s12903-023-03834-x.


Evaluation of Implant Stability According to Implant Placement Site and Duration in Elderly Patients: A Prospective Multi-Center Cohort Study.

Shim J, Kim M, An S, Kang E, Choi Y J Clin Med. 2023; 12(15).

PMID: 37568490 PMC: 10419881. DOI: 10.3390/jcm12155087.


Relevant Aspects of the Dental Implant Design on the Insertion Torque, Resonance Frequency Analysis (RFA) and Micromobility: An In Vitro Study.

Herrero-Climent M, Falcao A, Tondela J, Brizuela A, Rios-Carrasco B, Gil J J Clin Med. 2023; 12(3).

PMID: 36769501 PMC: 9917810. DOI: 10.3390/jcm12030855.


References
1.
Lachmann S, Jager B, Axmann D, Gomez-Roman G, Groten M, Weber H . Resonance frequency analysis and damping capacity assessment. Part I: an in vitro study on measurement reliability and a method of comparison in the determination of primary dental implant stability. Clin Oral Implants Res. 2006; 17(1):75-9. DOI: 10.1111/j.1600-0501.2005.01173.x. View

2.
Lages F, Douglas-de Oliveira D, Costa F . Relationship between implant stability measurements obtained by insertion torque and resonance frequency analysis: A systematic review. Clin Implant Dent Relat Res. 2017; 20(1):26-33. DOI: 10.1111/cid.12565. View

3.
Geckili O, Bilhan H, Cilingir A, Mumcu E, Bural C . A comparative in vitro evaluation of two different magnetic devices detecting the stability of osseo-integrated implants. J Periodontal Res. 2012; 47(4):508-13. DOI: 10.1111/j.1600-0765.2011.01462.x. View

4.
Lee J, Lim J, Lee J, Kim S, Koo K, Seol Y . Efficacy of sonic-powered toothbrushes for plaque removal in patients with peri-implant mucositis. J Periodontal Implant Sci. 2015; 45(2):56-61. PMC: 4415002. DOI: 10.5051/jpis.2015.45.2.56. View

5.
Turkyilmaz I, Sennerby L, Yilmaz B, Bilecenoglu B, Ozbek E . Influence of defect depth on resonance frequency analysis and insertion torque values for implants placed in fresh extraction sockets: a human cadaver study. Clin Implant Dent Relat Res. 2008; 11(1):52-8. DOI: 10.1111/j.1708-8208.2008.00095.x. View