Safety and Efficacy of a Novel Pneumatically Driven Extracorporeal Membrane Oxygenation Device

Overview
Journal Ann Thorac Surg
Publisher Elsevier
Date 2020 Mar 3
PMID 32119858
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Extracorporeal membrane oxygenation (ECMO) is rapidly becoming a mainstream technology for lung or heart/lung support. Current ECMO devices mostly consist of a power-driven centrifugal pump and a dedicated oxygenator. We studied the safety and efficacy of a novel, fully pneumatically driven ECMO device, which could be used in both venovenous or venoarterial mode in an animal model.

Methods: Six healthy, awake sheep were treated with the Mobybox ECMO device (Hemovent, Aachen, Germany) over a 7-day period in a venovenous mode. Gas exchange, coagulation parameters, and safety were assessed.

Results: Using a blood flow rate of 2 L/min and a low sweep gas flow rate of 0.3 L/min, the PCO ranged from 38 to 44 mm Hg pre oxygenator and dropped to 32 to 36 mm Hg post oxygenator, whereas the PaO post oxygenator increased to 600 mm Hg. Higher levels of sweep gas flow resulted in cessation of spontaneous breathing in some animals, consistent with high-efficiency carbon dioxide removal; thus, the sweep gas flow rate was maintained at a low level. Platelets dropped from 177 ± 53/μL to 107 ± 28/μL on day 2, while returning to baseline by day 7 (180 ± 51/μL). Plasma-free hemoglobin remained low (2-9 mg/dL), whereas fibrinogen slightly increased, and then remained stable throughout the period. Neither the pump nor the oxygenator showed any visible clotting after 7 days.

Conclusions: The pneumatically driven ECMO device provided excellent safety and physiologic efficacy in a 7-day sheep experiment without visible clotting, hemolysis, or sustained reductions in fibrinogen or platelets.

Citing Articles

Five-day Wearable Respiratory Support With a Novel Ambulatory Pulmonary Assist System in an Awake Ovine Model.

Hong Y, Shin S, Nasim U, Scala H, Roberts K, Potchernikov A Transplantation. 2024; 109(3):467-475.

PMID: 39160653 PMC: 11839337. DOI: 10.1097/TP.0000000000005165.


Progress towards permanent respiratory support.

Shin S, Nasim U, OConnor H, Hong Y Curr Opin Organ Transplant. 2024; 29(5):349-356.

PMID: 38990111 PMC: 11488683. DOI: 10.1097/MOT.0000000000001163.


Hemocompatibility Evaluation of a Novel Ambulatory Pulmonary Assist System Using a Lightweight Axial-Flow Pump.

Hong Y, Shin S, Nasim U, Roberts K, Potchernikov A, Liu K ASAIO J. 2024; 70(11):971-978.

PMID: 38728744 PMC: 11527577. DOI: 10.1097/MAT.0000000000002227.


Has Extracorporeal Gas Exchange Performance Reached Its Peak?.

Mouzakis F, Kashefi A, Hima F, Mottaghy K, Spillner J Membranes (Basel). 2024; 14(3).

PMID: 38535287 PMC: 10971993. DOI: 10.3390/membranes14030068.


Novel Size-Variable Dedicated Rodent Oxygenator for ECLS Animal Models-Introduction of the "RatOx" Oxygenator and Preliminary In Vitro Results.

Strudthoff L, Focke J, Hesselmann F, Kaesler A, Costa A, Schlanstein P Micromachines (Basel). 2023; 14(4).

PMID: 37421033 PMC: 10144398. DOI: 10.3390/mi14040800.