Flagging Performance of Sysmex XN-10 Haematology Analyser for Malaria Detection
Overview
Authors
Affiliations
Aim: The aim was to assess the flagging performance of Sysmex XN-10 haematology analyser for malaria detection through the parasitic red blood cell ('pRBC') alarm.
Methods: We retrospectively studied 584 blood samples performed on the Sysmex XN-10 analyser that were tested for malaria. Sensitivity, specificity, positive and negative predictive values, and prevalence were established for the pRBC alarm.
Results: Sensitivity, specificity, and positive and negative predictive values for the pRBC flag were 7.8%, 100%, 100% and 87.7%, respectively. The prevalence of pRBC flag of 0.026% in the overall population was significantly different from the prevalence of 1.027% in the population tested for malaria.
Conclusions: Considering the excellent specificity and the low prevalence of the flag in the overall population, we suggest, in case of the presence of pRBC flag, the implementation of a rapid review of the blood smear looking for , mostly if the patient had fever and had not been tested for malaria.
Rapid Detection of by the Hematology Analyzer for Population Screening.
Khodaiji S, Sehgal K, Sethi M, Mansukhani D Diagnostics (Basel). 2023; 13(22).
PMID: 37998533 PMC: 10670845. DOI: 10.3390/diagnostics13223397.
Procalcitonin as a Potential Biomarker in the Study of Babesiosis Caused by .
Lum M, Gauvin C, Pham S, Papamanoli A, Spitzer E, Kalogeropoulos A Pathogens. 2022; 11(11).
PMID: 36364974 PMC: 9698394. DOI: 10.3390/pathogens11111222.
Picot S, Perpoint T, Chidiac C, Sigal A, Javouhey E, Gillet Y Parasite. 2022; 29:31.
PMID: 35638753 PMC: 9153516. DOI: 10.1051/parasite/2022031.
Rastogi S, Nair S, Murugan P, Sukumar A, Mammen J, Mullai S Pract Lab Med. 2022; 30:e00275.
PMID: 35601458 PMC: 9117814. DOI: 10.1016/j.plabm.2022.e00275.