PERIPHERAL SURFACE DOSE FROM A LINEAR ACCELERATOR: RADIOCHROMIC FILM EXPERIMENTAL MEASUREMENTS OF FLATTENING FILTER FREE VERSUS FLATTENED BEAMS

Overview
Date 2020 Jan 11
PMID 31922571
Citations 1
Authors
Affiliations
Soon will be listed here.
Abstract

There is a growing interest in the use of flattening filter free (FFF) beams due to the shorter treatment times. The reduction of head scatter suggests a better radiation protection to radiotherapy patients, considering the expected decrease in peripheral surface dose (PSD). In this work, PSD of flattened (FF) and FFF-photon beams was compared. A radiochromic film calibration method to reduce energy dependence was used. PSD was measured at distances from 2 to 50 cm to the field border for different square field sizes, modifying relevant clinical parameters. Also, clinical breast and prostate stereotactic body radiotherapy (SBRT) plans were studied. For square beams, FFF PSD is lower compared with FF PSD (differences ranging from 3 to 64%) and 10 MV FFF yields to the lowest value, for distances greater than 5 cm. For SBRT plans, near and far away from the field border, there is a reduction of PSD for FFF-beams, but the behavior at intermediate distances should be checked depending on the case.