Detection of Peri-implant Bone Defects Using Cone-beam Computed Tomography and Digital Periapical Radiography with Parallel and Oblique Projection

Overview
Specialty Radiology
Date 2020 Jan 10
PMID 31915611
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To compare the diagnostic accuracy of cone-beam computed tomography (CBCT) with that of parallel (PPA) and oblique projected periapical (OPA) radiography for the detection of different types of peri-implant bone defects.

Materials And Methods: Forty implants inserted into bovine rib blocks were used. Thirty had standardized bone defects (10 each of angular, fenestration, and dehiscence defects), and 10 were defect-free controls. CBCT, PPA, and OPA images of the samples were acquired. The images were evaluated twice by each of 2 blinded observers regarding the presence or absence and the type of the defects. The area under the receiver operating characteristic curve (AUC), sensitivity, and specificity were determined for each radiographic technique. The 3 modalities were compared using the Fisher exact and chi-square tests, with <0.05 considered as statistical significance.

Results: High inter-examiner reliability was observed for the 3 techniques. Angular defects were detected with high sensitivity and specificity by all 3 modalities. CBCT and OPA showed similar AUC and sensitivity in the detection of fenestration defects. In the identification of dehiscence defects, CBCT showed the highest sensitivity, followed by OPA and PPA, respectively. CBCT and OPA had a significantly greater ability than PPA to detect fenestration and dehiscence defects (<0.05).

Conclusion: The application of OPA radiography in addition to routine PPA imaging as a radiographic follow-up method for dental implantation greatly enhances the visualization of fenestration and dehiscence defects. CBCT properly depicted all defect types studied, but it involves a relatively high dose of radiation and cost.

Citing Articles

Challenges in detecting various peri-implant bone defects on modified intraoral oblique radiographic projections: evaluation of an artificial mandibular model.

Oshima T, Asaumi R, Ogura S, Kawai T Oral Radiol. 2025; 41(2):248-259.

PMID: 39808239 DOI: 10.1007/s11282-024-00798-8.


Evaluation of the accuracy of cone beam computed tomography (CBCT) in the detection of peri-implant fenestration.

Gholampour A, Mollaei M, Ehsani H, Ghobadi F, Hosseinnataj A, Yazdani M BMC Oral Health. 2024; 24(1):922.

PMID: 39123157 PMC: 11316301. DOI: 10.1186/s12903-024-04674-z.


Diagnostic accuracy of cone-beam computed tomography with modified grayscale range for detection of buccal cortical plate defects adjacent to dental implants.

Yaghoobi S, Dalili Kajan Z, Khosravifard N, Khalighi Sigaroudi A, Jouybari R Dent Res J (Isfahan). 2023; 20:90.

PMID: 37810449 PMC: 10557990.


Analysis of peri-implant bone defects by using cone beam computed tomography (CBCT): an integrative review.

Costa J, Mendes J, Salazar F, Pacheco J, Rompante P, Camara M Oral Radiol. 2023; 39(3):455-466.

PMID: 37058184 PMC: 10244286. DOI: 10.1007/s11282-023-00683-w.


Efficacy of low-dose cone beam computed tomography and metal artifact reduction tool for assessment of peri-implant bone defects: an in vitro study.

Nomier A, Gaweesh Y, Taalab M, El Sadat S BMC Oral Health. 2022; 22(1):615.

PMID: 36528573 PMC: 9759909. DOI: 10.1186/s12903-022-02663-8.


References
1.
Mengel R, Kruse B, Flores-de-Jacoby L . Digital volume tomography in the diagnosis of peri-implant defects: an in vitro study on native pig mandibles. J Periodontol. 2006; 77(7):1234-41. DOI: 10.1902/jop.2006.050424. View

2.
Bagis N, Kolsuz M, Kursun S, Orhan K . Comparison of intraoral radiography and cone-beam computed tomography for the detection of periodontal defects: an in vitro study. BMC Oral Health. 2015; 15:64. PMC: 4446848. DOI: 10.1186/s12903-015-0046-2. View

3.
Ritter L, Elger M, Rothamel D, Fienitz T, Zinser M, Schwarz F . Accuracy of peri-implant bone evaluation using cone beam CT, digital intra-oral radiographs and histology. Dentomaxillofac Radiol. 2014; 43(6):20130088. PMC: 4141670. DOI: 10.1259/dmfr.20130088. View

4.
Ding Q, Zhang L, Geraets W, Wu W, Zhou Y, Wismeijer D . Association Between Peri-implant Bone Morphology and Marginal Bone Loss: A Retrospective Study on Implant-Supported Mandibular Overdentures. Int J Oral Maxillofac Implants. 2016; 32(1):147-155. DOI: 10.11607/jomi.4922. View

5.
Liedke G, Spin-Neto R, da Silveira H, Schropp L, Stavropoulos A, Wenzel A . Factors affecting the possibility to detect buccal bone condition around dental implants using cone beam computed tomography. Clin Oral Implants Res. 2016; 28(9):1082-1088. DOI: 10.1111/clr.12921. View