Nd:YAP Laser in the Treatment of Dentinal Hypersensitivity: An Ex Vivo Study

Overview
Specialty Biology
Date 2019 Dec 31
PMID 31887638
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: The aims of this ex vivo study were to evaluate the effectiveness of the Nd:YAP laser in the treatment of dentin hypersensitivity, to compare the temperature rise during laser irradiation at three different dentine thicknesses, and to analyse the composition of the dentine-lased surface.

Methods: A total of 33 teeth were used in this study. For scanning electron microscopy (SEM) observation, 24 teeth were transversely sectioned and divided into 4 groups: group A was irrigated with EDTA; group B was irradiated by Nd:YAP laser with 180 mJ energy/per pulse, 0.9 W average power, and 5 Hz frequency (power density [PD] = 229 W/cm); group C was irradiated by Nd:YAP laser with 280 mJ energy/pulse, 1.4 W average power, and 5 Hz frequency (PD = 356 W/cm); and group D was irradiated by Nd:YAP with 360 mJ energy/pulse, 1.8 W average power, and 5 Hz frequency (PD = 458 W/cm). Energy-dispersive spectroscopy (EDS) analysis was performed on the same teeth evaluated for SEM observations. For temperature increase evaluation performed with thermocouples, 9 teeth were transversely sectioned at 3 different thicknesses (3 for each group) of 1, 2, and 3 mm.

Results: Statistical analysis showed significant changes in the diameter of the dentinal tubule orifices among all groups; EDS did not show modification of the Ca/P ratio. Temperature increase under irradiation exceeded 5.5 °C only in the group D samples.

Conclusions: This ex vivo study, based on temperature recording, SEM observation, and EDS analysis, demonstrated that Nd:YAP laser at a PD of 356 W/cm, corresponding to an average power of 1.4 W, defines the best treatment for dentine hypersensitivity in terms of compromise between efficacy of the treatment and safety of the pulp.

Citing Articles

The effect of lasers in occlusion of dentinal tubules and reducing dentinal hypersensitivity, a scoping review.

Behniafar B, Noori F, Chiniforoush N, Raee A BMC Oral Health. 2024; 24(1):1407.

PMID: 39563326 PMC: 11575069. DOI: 10.1186/s12903-024-05182-w.


Evaluation of dentinal tubule occlusion and pulp tissue response after using 980-nm diode laser for dentin hypersensitivity treatment.

Meng Y, Huang F, Wang S, Huang X, Lu Y, Li Y Clin Oral Investig. 2023; 27(8):4843-4854.

PMID: 37382717 DOI: 10.1007/s00784-023-05114-y.


Current Applications and Future Directions of Lasers in Endodontics: A Narrative Review.

Huang Q, Li Z, Lyu P, Zhou X, Fan Y Bioengineering (Basel). 2023; 10(3).

PMID: 36978686 PMC: 10044917. DOI: 10.3390/bioengineering10030296.


Effects of Low-Level Laser Therapy on Dentin Hypersensitivity in Periodontally Compromised Patients Undergoing Orthodontic Treatment: A Randomised Controlled Trial.

Shan Z, Ren C, Gu M, Lin Y, Sum F, McGrath C J Clin Med. 2023; 12(4).

PMID: 36835953 PMC: 9961942. DOI: 10.3390/jcm12041419.


Twelve-Month Follow-Up of Different Dentinal Hypersensitivity Treatments by Photobiomodulation Therapy, Nd:YAG and Nd:YAP Lasers.

Nammour S, El Mobadder M, Namour M, Brugnera Junior A, Zanin F, Brugnera A Life (Basel). 2022; 12(12).

PMID: 36556361 PMC: 9784673. DOI: 10.3390/life12121996.