Algorithm-based Artifact Reduction in Patients with Arms-down Positioning in Computed Tomography

Overview
Journal Phys Med
Date 2019 Dec 13
PMID 31830632
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Arm-artifact, a type of streak artifact frequently observed in computed tomography (CT) images obtained at arms-down positioning in polytrauma patients, is known to degrade image quality. This study aimed to develop a novel arm-artifact reduction algorithm (AAR) applied to projection data.

Methods: A phantom resembling an adult abdomen with two arms was scanned using a 16-row CT scanner. The projection data were processed by AAR, and CT images were reconstructed. The artifact reduction for the same phantom was compared with that achieved by two latest iterative reconstruction (IR) techniques (IR1 and IR2) using a normalized artifact index (nAI) at two locations (ventral and dorsal side). Image blurring as a processing side effect was compared with IR2 of the model-based IR using a plastic needle phantom. Additionally, the projection data of two clinical cases were processed using AAR, and the image noise was evaluated.

Results: AAR and IR2 significantly reduced nAI by 87.5% and 74.0%, respectively at the ventral side and 84.2% and 69.6%, respectively, at the dorsal side compared with each filtered back projection (P < 0.01), whereas IR1 did not. The proposed algorithm mostly maintained the original spatial resolution, compared with IR2, which yielded apparent image blurring. The image noise in the clinical cases was also reduced significantly (P < 0.01).

Conclusions: AAR was more effective and superior than the latest IR techniques and is expected to improve the image quality of polytrauma CT imaging with arms-down positioning.

Citing Articles

[Deep Learning Reconstruction Algorithm Combined With Smart Metal Artifact Reduction Technique Improves Image Quality of Upper Abdominal CT in Critically Ill Patients].

Pan Y, Yao X, Gao R, Xie W, Xia C, Li Z Sichuan Da Xue Xue Bao Yi Xue Ban. 2025; 55(6):1403-1409.

PMID: 39990832 PMC: 11839371. DOI: 10.12182/20241160102.


An image-based metal artifact reduction technique utilizing forward projection in computed tomography.

Ichikawa K, Kawashima H, Takata T Radiol Phys Technol. 2024; 17(2):402-411.

PMID: 38546970 PMC: 11128408. DOI: 10.1007/s12194-024-00790-1.


PWLS-PR: low-dose computed tomography image reconstruction using a patch-based regularization method based on the penalized weighted least squares total variation approach.

Fu J, Feng F, Quan H, Wan Q, Chen Z, Liu X Quant Imaging Med Surg. 2021; 11(6):2541-2559.

PMID: 34079722 PMC: 8107320. DOI: 10.21037/qims-20-963.