Primary, Scatter, and Penetration Characterizations of Parallel-hole and Pinhole Collimators for I-123 SPECT

Overview
Journal Phys Med Biol
Publisher IOP Publishing
Date 2019 Nov 21
PMID 31746783
Citations 9
Authors
Affiliations
Soon will be listed here.
Abstract

Multi-pinhole (MPH) collimators are known to provide better trade-off between sensitivity and resolution for preclinical, as well as for smaller regions in clinical SPECT imaging compared to conventional collimators. In addition to this geometric advantage, MPH plates typically offer better stopping power for penetration than the conventional collimators, which is especially relevant for I-123 imaging. The I-123 emits a series of high-energy (>300 keV, ~2.5% abundance) gamma photons in addition to the primary emission (159 keV, 83% abundance). Despite their low abundance, high-energy photons penetrate through a low-energy parallel-hole (LEHR) collimator much more readily than the 159 keV photons, resulting in large downscatter in the photopeak window. In this work, we investigate the primary, scatter, and penetration characteristics of a single pinhole collimator that is commonly used for I-123 thyroid imaging and our two MPH collimators designed for I-123 DaTscan imaging for Parkinson's Disease, in comparison to three different parallel-hole collimators through a series of experiments and Monte Carlo simulations. The simulations of a point source and a digital human phantom with DaTscan activity distribution showed that our MPH collimators provide superior count performance in terms of high primary counts, low penetration, and low scatter counts compared to the parallel-hole and single pinhole collimators. For example, total scatter, multiple scatter, and collimator penetration events for the LEHR were 2.5, 7.6 and 14 times more than that of MPH within the 15% photopeak window. The total scatter fraction for LEHR was 56% where the largest contribution came from the high-energy scatter from the back compartments (31%). For the same energy window, the total scatter for MPH was 21% with only 1% scatter from the back compartments. We therefore anticipate that using MPH collimators, higher quality reconstructions can be obtained in a substantially shorter acquisition time for I-123 DaTscan and thyroid imaging.

Citing Articles

Evaluation of Improved Imaging Properties with Tungsten-Based Parallel-Hole Collimators: A Monte Carlo Study.

Islamian J, Ljungberg M World J Nucl Med. 2024; 23(2):95-102.

PMID: 38933066 PMC: 11199031. DOI: 10.1055/s-0044-1786165.


Multi-pinhole collimator design in different numbers of projections for brain SPECT.

Huang W, Mok G Front Med (Lausanne). 2023; 10:1211726.

PMID: 37841005 PMC: 10568733. DOI: 10.3389/fmed.2023.1211726.


Mesh modeling of system geometry and anatomy phantoms for realistic GATE simulations and their inclusion in SPECT reconstruction.

Auer B, Konik A, Fromme T, De Beenhouwer J, Kalluri K, Lindsay C Phys Med Biol. 2023; 68(7).

PMID: 36808915 PMC: 10073298. DOI: 10.1088/1361-6560/acbde2.


Machine learning-based prediction of conversion coefficients for I-123 metaiodobenzylguanidine heart-to-mediastinum ratio.

Okuda K, Nakajima K, Kitamura C, Ljungberg M, Hosoya T, Kirihara Y J Nucl Cardiol. 2023; 30(4):1630-1641.

PMID: 36740650 PMC: 10372132. DOI: 10.1007/s12350-023-03198-3.


Improved Performance of a Multipinhole SPECT for DAT Imaging by Increasing Number of Pinholes at the Expense of Increased Multiplexing.

Konik A, Zeraatkar N, Kalluri K, Auer B, Fromme T, He Y IEEE Trans Radiat Plasma Med Sci. 2021; 5(6):817-825.

PMID: 34746540 PMC: 8568304. DOI: 10.1109/trpms.2020.3035626.


References
1.
Bowen J, Huang Q, Ellin J, Lee T, Shrestha U, Gullberg G . Design and performance evaluation of a 20-aperture multipinhole collimator for myocardial perfusion imaging applications. Phys Med Biol. 2013; 58(20):7209-26. PMC: 3855225. DOI: 10.1088/0031-9155/58/20/7209. View

2.
Sydoff M, Lizana H, Mattsson S, Stabin M, Leide-Svegborn S . Determination of the biodistribution and dosimetry of ¹²³I-FP-CIT in male patients with suspected Parkinsonism or Lewy body dementia using planar and combined planar and SPECT/CT imaging. Appl Radiat Isot. 2013; 82:300-7. DOI: 10.1016/j.apradiso.2013.09.006. View

3.
van Roosmalen J, Goorden M, Beekman F . Molecular breast tomosynthesis with scanning focus multi-pinhole cameras. Phys Med Biol. 2016; 61(15):5508-28. DOI: 10.1088/0031-9155/61/15/5508. View

4.
Rault E, Staelens S, Van Holen R, De Beenhouwer J, Vandenberghe S . Accurate Monte Carlo modelling of the back compartments of SPECT cameras. Phys Med Biol. 2010; 56(1):87-104. DOI: 10.1088/0031-9155/56/1/006. View

5.
Darcourt J, Booij J, Tatsch K, Varrone A, Vander Borght T, Kapucu O . EANM procedure guidelines for brain neurotransmission SPECT using (123)I-labelled dopamine transporter ligands, version 2. Eur J Nucl Med Mol Imaging. 2009; 37(2):443-50. DOI: 10.1007/s00259-009-1267-x. View