Portable Accelerometer-Based Navigation System for Cup Placement of Total Hip Arthroplasty: A Prospective, Randomized, Controlled Study

Overview
Journal J Arthroplasty
Specialty Orthopedics
Date 2019 Sep 30
PMID 31563396
Citations 29
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Malposition of the acetabular component during total hip arthroplasty (THA) is associated with increased risk of dislocation, reduced range of motion, and accelerated wear. The purpose of this study is to compare cup positioning with a portable, accelerometer-based hip navigation system and conventional surgical technique.

Methods: In a prospective, randomized, clinical study, cups were implanted with a portable, accelerometer-based hip navigation system (navigation group; n = 55) or conventional technique (conventional group; n = 55). THA was conducted in the lateral position and through posterior approach. The cup position was determined postoperatively on pelvic radiograph and computed tomography scans.

Results: An average cup abduction of 39.2° ± 4.6° (range, 27° to 50°) and an average cup anteversion of 14.6° ± 6.1° (range, 1° to 27.5°) were found in the navigation group, and an average cup abduction of 42.9° ± 8.0° (range, 23° to 73°) and an average cup anteversion of 11.6° ± 7.7° (range, -12.1° to 25°) in the conventional group. A smaller variation in the navigation group was indicated for cup abduction (P = .001). The deviations from the target cup position were significantly lower in the navigation group (P = .001, .016). While only 37 of 55 cups in the conventional group were inside the Lewinnek safe zone, 51 of 55 cups in the navigation group were placed inside this safe zone (P = .006). The navigation procedure took a mean of 10 minutes longer than the conventional technique.

Conclusion: Use of the portable, accelerometer-based hip navigation system can improve cup positioning in THA.

Citing Articles

Impact of the difference between preoperative planning and intraoperative femoral neck osteotomy level on stem alignment.

Ueno S, Iwakiri K, Ohta Y, Minoda Y, Kobayashi A, Nakamura H Arch Orthop Trauma Surg. 2024; 145(1):24.

PMID: 39666028 DOI: 10.1007/s00402-024-05656-0.


Total Hip Arthroplasty Using the Conjoined Tendon-Preserving Posterior Approach: The Modified Posterior Approach with a Minimum of 2-Year Follow-Up.

Ozawa Y, Osawa Y, Seki T, Takegami Y, Iida H, Funahashi H Orthop Surg. 2024; 17(1):53-61.

PMID: 39600103 PMC: 11735358. DOI: 10.1111/os.14194.


Accuracy of Cup Placement Angle, Leg Lengthening, and Offset Measurement Using an AR-Based Portable Navigation System: Validation in Supine and Lateral Decubitus Positions for Total Hip Arthroplasty.

Ozaki Y, Yamamoto T, Kimura S, Kasai T, Niki R, Niki H Medicina (Kaunas). 2024; 60(10).

PMID: 39459508 PMC: 11509153. DOI: 10.3390/medicina60101721.


Accelerometer-based portable navigation system shows no superior accuracy over pelvic alignment guide for acetabular cup placement in total hip arthroplasty in lateral decubitus position.

Okamoto M, Takegami Y, Osawa Y, Tanaka S, Nobori T, Imagama S Arch Orthop Trauma Surg. 2024; 144(11):4849-4856.

PMID: 39347968 DOI: 10.1007/s00402-024-05571-4.


A Novel Device for Intraoperative Measurement of Stem Anteversion Angle in Total Hip Arthroplasty.

Iwakiri K, Ohta Y, Ueno S, Minoda Y, Kobayashi A, Nakamura H Arthroplast Today. 2024; 28:101458.

PMID: 39100413 PMC: 11295631. DOI: 10.1016/j.artd.2024.101458.