Abstract

Objectives: Preclinical data showed poly(methyl methacrylate) (PMMA) loaded with microsilver to be effective against a variety of bacteria. The purpose of this study was to assess patient safety of PMMA spacers with microsilver in prosthetic hip infections in a prospective cohort study.

Methods: A total of 12 patients with prosthetic hip infections were included for a three-stage revision procedure. All patients received either a gentamicin-PMMA spacer (80 g to 160 g PMMA depending on hip joint dimension) with additional loading of 1% (w/w) of microsilver (0.8 g to 1.6 g per spacer) at surgery 1 followed by a gentamicin-PMMA spacer without microsilver at surgery 2 or vice versa. Implantation of the revision prosthesis was carried out at surgery 3.

Results: In total, 11 of the 12 patients completed the study. No argyria or considerable differences in laboratory parameters were detected. Silver blood concentrations were below or around the detection limit of 1 ppb in ten of the 11 patients. A maximum of 5.6 ppb at 48 hours after implantation of the silver spacer, which is below the recommended maximum level of 10 ppb, was found in one patient. No silver was detected in the urine. Drainage fluids showed concentrations between 16.1 ppb and 23.3 ppb at 12 hours after implantation of the silver spacers, and between 16.8 ppb to 25.1 ppb at 48 hours after implantation. Pathohistological assessment of the periprosthetic membrane did not reveal any differences between the two groups.

Conclusion: Microsilver-loaded gentamicin-PMMA spacers showed good biocompatibility and the broad antimicrobial activity warrants further clinical research to assess its effectivity in reducing infection rates in prosthetic joint infection.: V. Alt, M. Rupp, K. Lemberger, T. Bechert, T. Konradt, P. Steinrücke, R. Schnettler, S. Söder, R. Ascherl. Safety assessment of microsilver-loaded poly(methyl methacrylate) (PMMA) cement spacers in patients with prosthetic hip infections: Results of a prospective cohort study. 2019;8:387-396. DOI: 10.1302/2046-3758.88.BJR-2018-0270.R1.

Citing Articles

What to Know about Antimicrobial Coatings in Arthroplasty: A Narrative Review.

Onorato F, Masoni V, Gagliardi L, Comba L, Rivera F Medicina (Kaunas). 2024; 60(4).

PMID: 38674220 PMC: 11052078. DOI: 10.3390/medicina60040574.


Physical, Mechanical, and Biological Properties of PMMA-Based Composite Bone Cement Containing Silver-Doped Bioactive and Antibacterial Glass Particles with Different Particles Sizes.

Miola M, Lucchetta G, Verne E Materials (Basel). 2023; 16(13).

PMID: 37444813 PMC: 10342656. DOI: 10.3390/ma16134499.


The sternum reconstruction: Present and future perspectives.

Aramini B, Masciale V, Radaelli L, Sgarzani R, Dominici M, Stella F Front Oncol. 2022; 12:975603.

PMID: 36387077 PMC: 9649912. DOI: 10.3389/fonc.2022.975603.


Trabecular Titanium for Orthopedic Applications: Balancing Antimicrobial with Osteoconductive Properties by Varying Silver Contents.

Diez-Escudero A, Carlsson E, Andersson B, Jarhult J, Hailer N ACS Appl Mater Interfaces. 2022; 14(37):41751-41763.

PMID: 36069272 PMC: 9501801. DOI: 10.1021/acsami.2c11139.


Antimicrobial rationing in orthopaedic surgery.

Tsang S, Simpson A Bone Joint Res. 2020; 9(12):870-872.

PMID: 33307769 PMC: 9021903. DOI: 10.1302/2046-3758.912.BJR-2020-0521.


References
1.
Neut D, van de Belt H, Stokroos I, van Horn J, van der Mei H, Busscher H . Biomaterial-associated infection of gentamicin-loaded PMMA beads in orthopaedic revision surgery. J Antimicrob Chemother. 2001; 47(6):885-91. DOI: 10.1093/jac/47.6.885. View

2.
Thomes B, Murray P, Bouchier-Hayes D . Development of resistant strains of Staphylococcus epidermidis on gentamicin-loaded bone cement in vivo. J Bone Joint Surg Br. 2002; 84(5):758-60. DOI: 10.1302/0301-620x.84b5.11907. View

3.
Alt V, Bechert T, Steinrucke P, Wagener M, Seidel P, Dingeldein E . An in vitro assessment of the antibacterial properties and cytotoxicity of nanoparticulate silver bone cement. Biomaterials. 2004; 25(18):4383-91. DOI: 10.1016/j.biomaterials.2003.10.078. View

4.
Gosheger G, Hardes J, Ahrens H, Streitburger A, Buerger H, Erren M . Silver-coated megaendoprostheses in a rabbit model--an analysis of the infection rate and toxicological side effects. Biomaterials. 2004; 25(24):5547-56. DOI: 10.1016/j.biomaterials.2004.01.008. View

5.
PERRELLI G, Piolatto G . Tentative reference values for gold, silver and platinum: literature data analysis. Sci Total Environ. 1992; 120(1-2):93-6. DOI: 10.1016/0048-9697(92)90219-i. View