Assessment of a Novel 3T MRI Compatible Cochlear Implant Magnet: Torque, Forces, Demagnetization, and Imaging

Overview
Journal Otol Neurotol
Date 2019 Sep 17
PMID 31524762
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Hypothesis: A novel cochlea implant (CI) device magnet providing alignment to the static field of a magnetic resonance imaging (MRI) will lead to reduced torque, longitudinal forces, and demagnetization effects. The image void and distortion will be comparable to those obtained with standard CI magnets.

Background: MRI investigations of CI users pose several challenges such as magnet dislocation, demagnetization, and may cause pain. The presence of a CI magnet within MRI field causes image distortions and may diminish the diagnostic value of an MRI procedure. Objective of this work is to evaluate magnetic forces and imaging properties of the novel CI magnet within 1.5 and 3T MRI.

Methods: Forces and torque of the novel CI magnet were measured in both 1.5 and 3T MRI and compared with the standard magnet in 1.5T. One cadaver head was implanted with the CI devices containing the novel and standard magnets in different configurations reflecting clinical scenarios and imaging properties were assessed and compared.

Results: In particular the torque has been reduced with the novel CI magnet in comparison to the standard one. Both CI magnets have not shown any signs of demagnetization. The image void and distortion was comparable between the two magnets for the main MRI clinical scanning protocols in 1.5T MRI.

Conclusions: The novel CI magnet is safe to use for MRI investigations of CI users in 3T MRI without a need for bandaging and has acceptable level of image artefacts.

Citing Articles

Multi-Magnet Cochlear Implant Technology and Magnetic Resonance Imaging: The Safety Issue.

Canzi P, Carlotto E, Zanoletti E, Frijns J, Borsetto D, Caruso A Audiol Res. 2024; 14(3):401-411.

PMID: 38804458 PMC: 11130805. DOI: 10.3390/audiolres14030034.


The Use of Portable, Very Low-field (0.064T) MRI to Image Cochlear Implants: Metallic Image Artifact in Comparison to Traditional, Stationary 3T MRI.

Munhall C, Roberts D, Labadie R Otol Neurotol Open. 2024; 4(1):e049.

PMID: 38533347 PMC: 10962874. DOI: 10.1097/ONO.0000000000000049.


A flexible omnidirectional rotating magnetic array for MRI-safe transdermal wireless energy harvesting through flexible electronics.

Zhou M, Mao S, Wu Z, Li Y, Yang Z, Liu X Sci Adv. 2023; 9(33):eadi5451.

PMID: 37585524 PMC: 10431719. DOI: 10.1126/sciadv.adi5451.


Risk of magnetic resonance imaging-induced magnet dislocation for different types of cochlear implants: a single-center retrospective study.

Rupp R, Balk M, Sievert M, Leibl V, Schleder S, Allner M J Otolaryngol Head Neck Surg. 2023; 52(1):28.

PMID: 37085925 PMC: 10122328. DOI: 10.1186/s40463-023-00633-w.


The role of cochlear implant positioning on MR imaging quality: a preclinical in vivo study with a novel implant magnet system.

Canzi P, Magnetto M, Simoncelli A, Manfrin M, Aprile F, Lafe E Eur Arch Otorhinolaryngol. 2021; 279(6):2889-2898.

PMID: 34370074 PMC: 9072450. DOI: 10.1007/s00405-021-07005-y.