» Articles » PMID: 31410911

Optimizing Acquisition and Fitting Conditions for H MR Spectroscopy Investigations in Global Brain Pathology

Overview
Journal NMR Biomed
Publisher Wiley
Date 2019 Aug 15
PMID 31410911
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To optimize acquisition and fitting conditions for nonfocal disease in terms of voxel size and use of individual coil element data. Increasing the voxel size yields a higher signal-to-noise ratio, but leads to larger linewidths and more artifacts. Several ways to improve the spectral quality for large voxels are exploited and the optimal use of individual coil signals investigated.

Methods: Ten human subjects were measured at 3 T using a 64-channel receive head coil with a semi-LASER localization sequence under optimized and deliberately mis-set field homogeneity. Eight different voxel sizes (8 to 99 cm ) were probed. Spectra were fitted either as weighted sums of the individual coil elements or simultaneously without summation. Eighteen metabolites were included in the fit model that also included the lineshapes from all coil elements as reflected in water reference data. Fitting errors for creatine, myo-Inositol and glutamate are reported as representative parameters to judge optimal acquisition and evaluation conditions.

Results: Minimal Cramér-Rao lower bounds and thus optimal acquisition conditions were found for a voxel size of ~ 70 cm for the representative upfield metabolites. Spectral quality in terms of lineshape and artifact appearance was determined to differ substantially between coil elements. Simultaneous fitting of spectra from individual coil elements instead of traditional fitting of a weighted sum spectrum reduced Cramer-Rao lower bounds by up to 17% for large voxel sizes.

Conclusion: The optimal voxel size for best precision in determined metabolite content is surprisingly large. Such an acquisition condition is most relevant for detection of low-concentration metabolites, like NAD or phenylalanine, but also for longitudinal studies where very small alterations in metabolite content are targeted. In addition, simultaneous fitting of single channel spectra enforcing lineshape and coil sensitivity information proved to be superior to traditional signal combination with subsequent fitting.

Citing Articles

Optimization of H-MRS methods for large-volume acquisition of low-concentration downfield resonances at 3 T and 7 T.

Wilson N, Elliott M, Nanga R, Swago S, Witschey W, Reddy R Magn Reson Med. 2024; 93(1):18-30.

PMID: 39250517 PMC: 11518639. DOI: 10.1002/mrm.30273.


Transient brain structure changes after high phenylalanine exposure in adults with phenylketonuria.

Muri R, Rummel C, McKinley R, Rebsamen M, Maissen-Abgottspon S, Kreis R Brain. 2024; 147(11):3863-3873.

PMID: 38723047 PMC: 11604053. DOI: 10.1093/brain/awae139.


Simultaneous multi-transient linear-combination modeling of MRS data improves uncertainty estimation.

Zollner H, Davies-Jenkins C, Simicic D, Tal A, Sulam J, Oeltzschner G Magn Reson Med. 2024; 92(3):916-925.

PMID: 38649977 PMC: 11209799. DOI: 10.1002/mrm.30110.


Optimization of H MR spectroscopy methods for large volume acquisition of low concentration downfield resonances at 3T and 7T.

Wilson N, Elliott M, Nanga R, Swago S, Witschey W, Reddy R medRxiv. 2024; .

PMID: 38645233 PMC: 11030301. DOI: 10.1101/2024.04.09.24305552.


Simultaneous multi-transient linear-combination modeling of MRS data improves uncertainty estimation.

Zollner H, Davies-Jenkins C, Simicic D, Tal A, Sulam J, Oeltzschner G bioRxiv. 2024; .

PMID: 38260650 PMC: 10802456. DOI: 10.1101/2023.11.01.565164.