[Evaluation of Bispectral Index Time Delay in Response to Anesthesia Induction: an Observational Study]
Overview
Affiliations
Background And Objectives: According to the manufacturer, the Bispectral Index (BIS) has a processing time delay of 5-10s. Studies addressing this have suggested longer delays. We evaluated the time delay in the Bispectral Index response.
Methods: Based on clinical data from 45 patients, using the difference between the predicted and the real BIS, calculated during a fixed 3minutes period after the moment the Bispectral Index dropped below 80 during the induction of general anesthesia with propofol and remifentanil.
Results: The difference between the predicted and the real BIS was in average 30.09±18.73s.
Conclusion: Our results may be another indication that the delay in BIS processing may be much longer than stated by the manufacture, a fact with clinical implications.
Chen L, Xie H, Huang X, Luo T, Guo J, Lin C Nan Fang Yi Ke Da Xue Xue Bao. 2025; 45(1):52-58.
PMID: 39819712 PMC: 11744284. DOI: 10.12122/j.issn.1673-4254.2025.01.07.
Singla D, Agrawal S, T K P, Adhikary A, Mangla M Turk J Anaesthesiol Reanim. 2024; 52(4):154-160.
PMID: 39287212 PMC: 11590694. DOI: 10.4274/TJAR.2024.241663.
Frelich M, Sklienka P, Romanova T, Nemcova S, Bilena M, Strakova H BMC Anesthesiol. 2024; 24(1):228.
PMID: 38982400 PMC: 11232140. DOI: 10.1186/s12871-024-02610-w.
Substance-dependent EEG during recovery from anesthesia and optimization of monitoring.
Lipp M, Schneider G, Kreuzer M, Pilge S J Clin Monit Comput. 2023; 38(3):603-612.
PMID: 38108943 PMC: 11164797. DOI: 10.1007/s10877-023-01103-4.
Leroy S, Major S, Bublitz V, Dreier J, Koch S Front Aging Neurosci. 2023; 14:1076393.
PMID: 36742202 PMC: 9889977. DOI: 10.3389/fnagi.2022.1076393.