An Inter-laboratory Comparison to Evaluate the Technical Performance of Rabies Diagnosis Lateral Flow Assays

Overview
Journal J Virol Methods
Specialty Microbiology
Date 2019 Jul 28
PMID 31351168
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

As in previous years, the European Union Reference Laboratory (EURL) for rabies organised in 2018 an Inter-laboratory trial (ILT) on rabies diagnosis. Contrarily to past years, the 2018 ILT did not aim to evaluate the performance of participating laboratories, but the technical performance of new rapid tests. Two lateral Flow Assays (LFA), namely the Anigen® and the CDIA™ Rabies Virus Antigen Rapid Test" (commercialized by Bionote and Creative Diagnostics Cie respectively), were evaluated together with the Fluorescent Antibody Test (FAT). One panel of virus samples (including RABV as well as EBLV1a, EBLV-1b, and EBLV2 strains) was sent to participating laboratories to compare results obtained with these different techniques. The study revealed that the FAT provided a good agreement toward expected results for both negative/positive samples (99.1%). The Anigen® test produced similar results to the FAT, with only one false negative result (0.5%) reported by all participants and a concordance of 100% for all but one sample demonstrating a good inter-laboratory reproducibility of the Anigen® batch. The CDIA™ test produced reproducible results for Rabies Virus (RABV) samples only. However, it hardly detected the Bokeloh Bat Lyssavirus (BBLV) and the European Bat Lyssaviruses types 1b and 2 (EBLV-1b and EBLV-2) in most laboratories resulting in a moderate inter-laboratory concordance (58.4%-82.7%) for these lyssaviruses. The two LFAs provided reliable and reproducible results on all RABV samples (100%) but lead to heterogeneous performances with other lyssaviruses leading to different levels of diagnostic/analytical sensitivity, specificity. The study confirmed that LFAs should be used with caution and that their validation are of upmost importance before any use in laboratories.

Citing Articles

Investigating the Impact That Diagnostic Screening with Lateral Flow Devices Had on the Rabies Surveillance Program in Zanzibar, Tanzania.

Mohd A, Coetzer A, Malan A, Scott T, Ramadhan R, Wright N Microorganisms. 2024; 12(7).

PMID: 39065083 PMC: 11279036. DOI: 10.3390/microorganisms12071314.


Comparison of 4 laboratory tests for the detection of bovine rabies viral infection in Paraguay: fluorescent antibody test, rapid detection test, histologic lesions, and RT-PCR.

Rodriguez M, Rodriguez R, Rodriguez A, Batista H, Samudio M, Cardozo W J Vet Diagn Invest. 2024; 36(4):522-528.

PMID: 38653733 PMC: 11185112. DOI: 10.1177/10406387241246712.


Recent updates on laboratory diagnosis of rabies.

Ashwini M, Pattanaik A, Mani R Indian J Med Res. 2024; 159(1):48-61.

PMID: 38376376 PMC: 10954107. DOI: 10.4103/ijmr.ijmr_131_23.


Evaluation of a real-time mobile PCR device (PCR 1100) for the detection of the rabies gene in field samples.

Demetria C, Kimitsuki K, Yahiro T, Saito N, Hashimoto T, Khan S Trop Med Health. 2023; 51(1):17.

PMID: 36932428 PMC: 10020757. DOI: 10.1186/s41182-023-00501-3.


Capacity Building Efforts for Rabies Diagnosis in Resource-Limited Countries in Sub-Saharan Africa: A Case Report of the Central Veterinary Laboratory in Benin (Parakou).

Djegui F, Gourlaouen M, Coetzer A, Adjin R, Tohozin R, Leopardi S Front Vet Sci. 2022; 8:769114.

PMID: 35118149 PMC: 8805029. DOI: 10.3389/fvets.2021.769114.