A Blood Flow Volume Linear Inversion Model Based on Electromagnetic Sensor for Predicting the Rate of Arterial Stenosis

Overview
Journal Sensors (Basel)
Publisher MDPI
Specialty Biotechnology
Date 2019 Jul 11
PMID 31288428
Citations 4
Authors
Affiliations
Soon will be listed here.
Abstract

This paper presents a mathematical model of measuring blood flow based on electromagnetic induction for predicting the rate of arterial stenosis. Firstly, an electrode sensor was used to collect the induced potential differences from human skin surface in a uniform magnetic field. Then, the inversion matrix was constructed by the weight function theory and finite element method. Next, the blood flow volume inversion model was constructed by combining the induction potential differences and inversion matrix. Finally, the rate of arterial stenosis was predicted based on mathematical relationship between blood flow and the area of arterial stenosis. To verify the accuracy of the model, a uniform magnetic field distribution of Helmholtz coil and a 3D geometric model of the ulnar artery of the forearm with different rates of stenosis were established in COMSOL, a finite element analysis software. Simulation results showed that the inversion model had high accuracy in the measurement of blood flow and the prediction of rate of stenosis, and is of great significance for the early diagnosis of arterial stenosis and other vessel diseases.

Citing Articles

Characterization and potential health benefits of millet flour and banana peel mixtures on rats fed with a high-fat diet.

Alshehri A, Younes N, Kamel R, Shawir S Heliyon. 2024; 10(20):e39424.

PMID: 39497975 PMC: 11532225. DOI: 10.1016/j.heliyon.2024.e39424.


[Research on inversion method of intravascular blood flow velocity based on convolutional neural network].

Wang Y, Yang D, Xu B, Zhang X, Wang X Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2022; 39(3):561-569.

PMID: 35788526 PMC: 10950770. DOI: 10.7507/1001-5515.202112038.


A Deep Neural Network Method for Arterial Blood Flow Profile Reconstruction.

Yang D, Wang Y, Xu B, Wang X, Liu Y, Cheng T Entropy (Basel). 2021; 23(9).

PMID: 34573739 PMC: 8467034. DOI: 10.3390/e23091114.


Application of Linear Gradient Magnetic Field in Arterial Profile Scanning Imaging.

Liu Y, Liu G, Yang D, Xu B Sensors (Basel). 2020; 20(16).

PMID: 32823785 PMC: 7472423. DOI: 10.3390/s20164547.

References
1.
Maeda K, Mies G, Olah L, Hossmann K . Quantitative measurement of local cerebral blood flow in the anesthetized mouse using intraperitoneal [14C]iodoantipyrine injection and final arterial heart blood sampling. J Cereb Blood Flow Metab. 2000; 20(1):10-4. DOI: 10.1097/00004647-200001000-00003. View

2.
Randoux B, Marro B, Koskas F, Duyme M, Sahel M, Zouaoui A . Carotid artery stenosis: prospective comparison of CT, three-dimensional gadolinium-enhanced MR, and conventional angiography. Radiology. 2001; 220(1):179-85. DOI: 10.1148/radiology.220.1.r01jl35179. View

3.
Delille J, Slanetz P, Yeh E, Kopans D, Garrido L . Breast cancer: regional blood flow and blood volume measured with magnetic susceptibility-based MR imaging--initial results. Radiology. 2002; 223(2):558-65. DOI: 10.1148/radiol.2232010428. View

4.
Aschenbach R, Esser D . [Magnetic resonance angiography of the head and neck area]. HNO. 2004; 52(1):77-89. DOI: 10.1007/s00106-003-0982-z. View

5.
Lui E, Steinman A, Cobbold R, Johnston K . Human factors as a source of error in peak Doppler velocity measurement. J Vasc Surg. 2005; 42(5):972-9. DOI: 10.1016/j.jvs.2005.07.014. View