Compare the Accuracy and Precision of Coulter LH780, Mindray BC-6000 Plus, and Sysmex XN-9000 with the International Reference Flow Cytometric Method in Platelet Counting

Overview
Journal PLoS One
Date 2019 May 25
PMID 31125378
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: The aim of this study is to evaluate the performance of different platelet counting methods (optical, impedance, fluorescence and hand counting) applied in different analysers by comparing with the international flow cytometric reference method (IRM).

Methods: A total of 333 blood samples from different subgroups (168 cases with thrombocytopenia, 136 cases with normal platelet counts and 29 cases with thrombocytosis) were tested. Regarding IRM as the gold standard, we compared the accuracy and precision of different platelet count methods; i.e. LH780 (impedance), BC-6000 Plus (optical (O) and impedance (I)), Sysmex XN-9000 (optical (O), impedance (I), fluorescence (F)), and hand counting.

Results: Sysmex XN-9000-F (r = 0.988) had the best correlation with IRM for thrombocytopenic samples; BC-6000 Plus-I (r = 0.966) was more relevant to IRM than any other method for samples with normal platelet counts. Correlation between Sysmex XN-9000-I (r = 0.960) and IRM was the highest among these methods for samples with thrombocytosis. For bias evaluation, the average bias of Sysmex XN-9000-F was -1.5 × 109/L (95% LA = -9.4 to +6.4) for samples with thrombocytopenia, compared with IRM. BC-6000 Plus-I had a small mean difference with IRM for samples with normal platelet counts or thrombocytosis. Moreover, all evaluated methods had acceptable sensitivity, specificity, and concordance rates as compared with IRM in the diagnosis of thrombocytopenia and thrombocytosis.

Conclusions: Platelet counting by Sysmex XN-9000-F is more accurate than other methods for thrombocytopenic samples. BC-6000 Plus-I has superior association and consistency for normal platelet counts. As for thrombocytosis patients, Sysmex XN-9000-I has the highest correlation with IRM while Sysmex XN-9000-O has the highest diagnosis efficacy.

Citing Articles

Application of image recognition technology in pathological diagnosis of blood smears.

Cheng W, Liu J, Wang C, Jiang R, Jiang M, Kong F Clin Exp Med. 2024; 24(1):181.

PMID: 39105953 PMC: 11303489. DOI: 10.1007/s10238-024-01379-z.


Performance evaluation between two automated biochemical analyzer systems: Roche Cobas 8000 and Mindray BS2000M.

Chen M, Qin S, Yang S, Chen H, Lu L, Qin X J Med Biochem. 2022; 41(3):306-315.

PMID: 36042904 PMC: 9375536. DOI: 10.5937/jomb0-34328.


Performance of Platelet Counting in Thrombocytopenic Samples: Comparison between Mindray BC-6800Plus and Sysmex XN-9000.

Kim H, Hur M, Lee G, Kim S, Moon H, Yun Y Diagnostics (Basel). 2022; 12(1).

PMID: 35054235 PMC: 8775070. DOI: 10.3390/diagnostics12010068.


Performance evaluation of optical platelet counting of BC-6000Plus automated hematology analyzer.

Hu X, Tang Y, Li M, Fu C, Deng Z, Tang W Ann Transl Med. 2021; 9(6):508.

PMID: 33850905 PMC: 8039664. DOI: 10.21037/atm-21-691.


Pseudothrombocytopenia-A Review on Causes, Occurrence and Clinical Implications.

Lardinois B, Favresse J, Chatelain B, Lippi G, Mullier F J Clin Med. 2021; 10(4).

PMID: 33557431 PMC: 7915523. DOI: 10.3390/jcm10040594.

References
1.
OShea C, Werre S, Dahlgren L . Comparison of platelet counting technologies in equine platelet concentrates. Vet Surg. 2014; 44(3):304-13. DOI: 10.1111/j.1532-950X.2014.12290.x. View

2.
Sora F, Iurlo A, Sica S, Latagliata R, Annunziata M, Galimberti S . Chronic myeloid leukaemia with extreme thrombocytosis at presentation: incidence, clinical findings and outcome. Br J Haematol. 2017; 181(2):267-270. DOI: 10.1111/bjh.14553. View

3.
Segal H, Briggs C, Kunka S, Casbard A, Harrison P, Machin S . Accuracy of platelet counting haematology analysers in severe thrombocytopenia and potential impact on platelet transfusion. Br J Haematol. 2005; 128(4):520-5. DOI: 10.1111/j.1365-2141.2004.05352.x. View

4.
Demirtas S, Karahan O, Yazici S, Guclu O, Caliskan A, Yavuz C . The relationship between complete blood count parameters and Fontaine's Stages in patients with peripheral arterial disease. Vascular. 2014; 22(6):427-31. DOI: 10.1177/1708538114522227. View

5.
Johannessen B, Haugen T, Scott C . Standardisation of platelet counting accuracy in blood banks by reference to an automated immunoplatelet procedure: comparative evaluation of Cell-Dyn CD4000 impedance and optical platelet counts. Transfus Apher Sci. 2002; 25(2):93-106. DOI: 10.1016/s1473-0502(01)00099-4. View