» Articles » PMID: 31055574

Single Session and Weekly Beta 2-Microglobulin Removal with Different Dialytic Procedures: Comparison Between High-Flux Standard Bicarbonate Hemodialysis, Post-Dilution Hemodiafiltration, Short Frequent Hemodialysis with NxStage Technology And...

Overview
Journal Blood Purif
Specialty Hematology
Date 2019 May 6
PMID 31055574
Citations 9
Authors
Affiliations
Soon will be listed here.
Abstract

Background: NxStage System One cycler (NSO) is a widespread system for home daily dialysis. Few data are available on the impact of this "low dialysate volumes system" on the removal rate of poorly diffusible, time-dependent solutes like β2-microglobulin (β2M).

Methods: Single-session and weekly balances of β2M were performed and compared in 12 patients on daily NSO, 13 patients on standard high-flux bicarbonate dialysis (BHD), 5 patients on standard post-dilution on line hemodiafiltration (HDF), and 13 patients on automated peritoneal dialysis (APD).

Results: Intradialytic fall of plasma water β2M levels (corrected for rebound) was 65.2 ± 2.6% in HDF, 49.8 ± 9.1% in BHD, and 32.3 ± 6.4% in NSO (p < 0.001 between all groups). Single treatment dialysate removal was much less in APD (19.4 ± 20.4 mg, p < 0.001) than in any extracorporeal technologies, and was less in NSO (126.2 ± 35.6 mg, p < 0.001) than in BHD (204.9 ± 53.4 mg) and HDF (181.9 ± 37.6 mg), with no differences between the latter 2; however weekly removal was higher in NSO (757.3 ± 213.7 mg, p < 0.04) than in BHD (614.8 ± 160.3 mg) and HDF (545.8 ± 112.8 mg). Extrapolated β2M adsorption to the membrane was negligible in BHD, 14.7 ± 9.5% of total removal in HDF and 18.3 ± 18.5% in NSO. Integration of single session data into a weekly efficiency indicator (K × t) showed total volume of plasma cleared in NSO (33.4 ± 7.7 L/week) to be higher than in BHD (26.9 ± 7.2 L/week, p < 0.01) and not different than in HDF (36.2 ± 4.7 L/week); it was negligible (3.2 ± 1.0) in APD.

Conclusions: Weekly β2M removal efficiency proved equal and highest in HDF and NSO (at a 6/week prescription), slightly lesser in BHD and lowest in APD.

Citing Articles

Effects of Medium Cut-Off Polyarylethersulfone and Polyvinylpyrrolidone Blend Membrane Dialyzers in Hemodialysis Patients: A Systematic Review and Meta-Analysis of Randomized Controlled Trials.

Hung Y, Lai T, Belmouaz M, Tu Y, Lai C, Lin S Membranes (Basel). 2022; 12(5).

PMID: 35629769 PMC: 9144787. DOI: 10.3390/membranes12050443.


Safety and Efficacy of Short Daily Hemodialysis with Physidia S System: Clinical Performance Assessment during the Training Period.

Fessi H, Szelag J, Courivaud C, Nicoud P, Aguilera D, Gilbert O J Clin Med. 2022; 11(8).

PMID: 35456216 PMC: 9031690. DOI: 10.3390/jcm11082123.


Association between serum β2-microglobulin and mortality in Japanese peritoneal dialysis patients: A cohort study.

Maruyama Y, Nakayama M, Abe M, Yokoo T, Minakuchi J, Nitta K PLoS One. 2022; 17(4):e0266882.

PMID: 35421178 PMC: 9009671. DOI: 10.1371/journal.pone.0266882.


High-volume hemodiafiltration decreases the pre-dialysis concentrations of indoxyl sulfate and p-cresyl sulfate compared to hemodialysis: a post-hoc analysis from the HDFit randomized controlled trial.

Lima J, Guedes M, Rodrigues S, Florido A, Moreno-Amaral A, Barra A J Nephrol. 2022; 35(5):1449-1456.

PMID: 35239175 DOI: 10.1007/s40620-022-01283-3.


Optical Method and Biochemical Source for the Assessment of the Middle-Molecule Uremic Toxin β2-Microglobulin in Spent Dialysate.

Paats J, Adoberg A, Arund J, Fridolin I, Lauri K, Leis L Toxins (Basel). 2021; 13(4).

PMID: 33807387 PMC: 8067283. DOI: 10.3390/toxins13040255.