» Articles » PMID: 31011608

Comparison Between Roche Integra 400 Plus and Abbott Architect Ci8200 in Alanine Aminotransferase Assay

Overview
Journal Pract Lab Med
Specialty Biochemistry
Date 2019 Apr 24
PMID 31011608
Citations 1
Authors
Affiliations
Soon will be listed here.
Abstract

Background: The aim of this work is to present the results of a comparative study between the ALT assay on Integra 400 plus Roche Diagnostic versus Architect ci8200 of Abbott Diagnostic.

Methods: A total of 200 patients hospitalized in the various departments of the university hospital Mohammed VI of Oujda were prospectively tested on two systems: Abbott Architect ci8200 and Roche Integra 400 plus. Both analyzers use the spectrophotometric technique by coupling the transamination reaction to the oxidation-reduction reaction at NAD. The agreement of the results between the different techniques was evaluated using the Bland-Altman difference diagram and the Passing-Bablok and Deming regression line.

Results: There was a high concordance between the two assays: the equation of the Passing-Bablok line is YArchitect ​= ​-0,5625 ​+ ​0,9917 XIntegra with a correlation coefficient r = 0.999. The Bland-Altman diagram shows that the mean bias between the two methods is of the order of 1.1 IU/L and the difference between the ALT measurements by Architect and Integra is in the range of -1.4 to 3.6.

Conclusion: Our study shows a high correlation of the ALT assay results between the architect ci8200 and Integra 400 plus.

Citing Articles

Performance evaluation of QuantStudio 1 plus real-time PCR instrument for clinical laboratory analysis: A proof-of-concept study.

Wang Z, Yi J, Yu Q, Liu Y, Zhang R, Zhang D Pract Lab Med. 2023; 36:e00330.

PMID: 37649547 PMC: 10462677. DOI: 10.1016/j.plabm.2023.e00330.

References
1.
Lawlor D, Sattar N, Davey Smith G, Ebrahim S . The associations of physical activity and adiposity with alanine aminotransferase and gamma-glutamyltransferase. Am J Epidemiol. 2005; 161(11):1081-8. DOI: 10.1093/aje/kwi125. View

2.
Fraser A, Ebrahim S, Davey Smith G, Lawlor D . A comparison of associations of alanine aminotransferase and gamma-glutamyltransferase with fasting glucose, fasting insulin, and glycated hemoglobin in women with and without diabetes. Hepatology. 2007; 46(1):158-65. DOI: 10.1002/hep.21667. View

3.
Goessling W, Massaro J, Vasan R, DAgostino Sr R, Ellison R, Fox C . Aminotransferase levels and 20-year risk of metabolic syndrome, diabetes, and cardiovascular disease. Gastroenterology. 2008; 135(6):1935-44, 1944.e1. PMC: 3039001. DOI: 10.1053/j.gastro.2008.09.018. View

4.
Ruhl C, Everhart J . Elevated serum alanine aminotransferase and gamma-glutamyltransferase and mortality in the United States population. Gastroenterology. 2008; 136(2):477-85.e11. DOI: 10.1053/j.gastro.2008.10.052. View

5.
Fraser A, Harris R, Sattar N, Ebrahim S, Davey Smith G, Lawlor D . Alanine aminotransferase, gamma-glutamyltransferase, and incident diabetes: the British Women's Heart and Health Study and meta-analysis. Diabetes Care. 2009; 32(4):741-50. PMC: 2660465. DOI: 10.2337/dc08-1870. View