Radiographic Assessment of Overlengthening of the MoPyC Radial Head Prosthesis: a Cadaveric Study

Overview
Date 2019 Apr 17
PMID 30989412
Citations 4
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: The aim of this study was to evaluate a radiographic measurement method for assessment of overlengthening of the MoPyC radial head prosthesis.

Materials And Methods: Seven cadaver specimens were studied in ten stages: native specimen (1), radial head resection (2), and implantation of the MoPyC radial head prostheses (Bioprofile, Tornier, Montbonnot-Saint-Martin, France) in four increasing length (correct length, overlengthening of 1.5 mm, 3 mm, and 5 mm) with an intact medial collateral ligament (MCL 3-6) and following transection of the MCL (7-10). The radiographic measurement method according to Athwal et al. was evaluated to detect overlengthening. Statistical analysis included calculation of the diagnostic accuracy of the radiographic method.

Results: The radiographic measurement method correctly determined the size of the radial head prosthesis within ± 1 mm in 224 of 336 scenarios (67%) and within ± 2 mm in 320 of 336 scenarios (95%). With a threshold value of ≥ 1 mm, the overall diagnostic sensitivity for detecting overlengthening when it was present and was 90% and the specificity was 79%. The sensitivity was higher with increasing size of the prosthesis: in cases with overlengthening of 1.5 mm, the sensitivity was 76%, with an overlengthening of 3 mm, the sensitivity was 95%, and with an overlengthening of 5 mm, the sensitivity was 100%.

Conclusion: The radiographic measurement method of Athwal et al. can be used to estimate and to diagnose the magnitude overlengthening of the MoPyC radial head prosthesis. However, the sensitivity is limited (76%) in cases with a small amount of overlengthening of 1.5 mm.

Citing Articles

Monteggia Equivalent Variant Type 1 in an Adult, its Management and Functional Outcome with Literature Review.

Mukherjee S, Parkaviyan R, Banerjee K J Orthop Case Rep. 2024; 14(10):163-168.

PMID: 39381270 PMC: 11458213. DOI: 10.13107/jocr.2024.v14.i10.4848.


Intraoperative visualization of the posterolateral ulnohumeral joint space is reliable to indicate overlengthening in radial head arthroplasty.

Rippke J, Mengis N, Benken S, Schneider M, Prescher A, Hackl M Arch Orthop Trauma Surg. 2023; 144(3):1047-1053.

PMID: 38114739 DOI: 10.1007/s00402-023-05154-9.


Fluoroscopy of the Elbow: A Cadaveric Study Defining New Standard Projections to Visualize Important Anatomical Landmarks.

Muller S, Adolfsson L, Baum C, Muller-Gerbl M, Muller A, Rikli D JB JS Open Access. 2021; 6(2).

PMID: 34056508 PMC: 8154458. DOI: 10.2106/JBJS.OA.20.00160.


Monteggia-like lesions in adults treated with radial head arthroplasty-mid-term follow-up of 27 cases.

Jung M, Groetzner-Schmidt C, Porschke F, Grutzner P, Guehring T, Schnetzke M J Orthop Surg Res. 2020; 15(1):5.

PMID: 31900241 PMC: 6942379. DOI: 10.1186/s13018-019-1540-z.