A Six-degree-of-freedom Robotic Motion System for Quality Assurance of Real-time Image-guided Radiotherapy

Overview
Journal Phys Med Biol
Publisher IOP Publishing
Date 2019 Apr 16
PMID 30986773
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

In this study we develop and characterise a six degree-of-freedom (6 DoF) robotic motion system for quality assurance of real-time image-guided radiotherapy techniques. The system consists of a commercially available robotic arm, an acrylic phantom with embedded Calypso markers, a custom base plate to mount the robot to the treatment couch, and control software implementing the appropriate sequence of transformations to reproduce measured tumour motion traces. The robotic motion system was evaluated in terms of the set-up and motion trace repeatability, static localization accuracy and dynamic localization accuracy. Four prostate, two liver and three lung motion traces, representing a range of tumor motion trajectories recorded in real patient treatments, were executed using the robotic motion system and compared with motion measurements from the clinical Calypso motion tracking system. System set-up and motion trace repeatability was better than 0.5 deg and 0.3 mm for rotation and translation, respectively. The static localization accuracy of the robotic motion system in the LR, SI and AP directions was 0.09 mm, 0.08 mm and 0.02 mm for translations, respectively, and 0.2°, 0.06° and 0.06° for rotations, respectively. The dynamic localization accuracy of the robotic motion system was  <0.2 mm and  <0.6° for translations and rotations, respectively. Thus, we have demonstrated the ability to accurately mimic rigid-body tumor motion using a robotically controlled phantom to provide precise geometric QA for advanced radiotherapy delivery approaches.

Citing Articles

Enhancing intrafraction position monitoring for prostate radiotherapy on a conventional linear accelerator: an optimization study.

Arumugam S Phys Eng Sci Med. 2025; .

PMID: 39992544 DOI: 10.1007/s13246-025-01527-z.


A Multimodal Point Cloud-Based Method for Tumor Localization in Robotic Ultrasound-Guided Radiotherapy.

Song L, Li Q, Sai Z, Sun K, Chen W, Zhang H Technol Cancer Res Treat. 2024; 23:15330338241273149.

PMID: 39155658 PMC: 11491879. DOI: 10.1177/15330338241273149.


The Remove-the-Mask Open-Source head and neck Surface-Guided radiation therapy system.

Ben Bouchta Y, Gardner M, Sengupta C, Johnson J, Keall P Phys Imaging Radiat Oncol. 2024; 29:100541.

PMID: 38327762 PMC: 10847032. DOI: 10.1016/j.phro.2024.100541.


Assessment of Collaborative Robot (Cobot)-Assisted Histotripsy for Venous Clot Ablation.

Bader K, Hendley S, Bollen V IEEE Trans Biomed Eng. 2020; 68(4):1220-1228.

PMID: 32915723 PMC: 8018710. DOI: 10.1109/TBME.2020.3023630.