Solid Phase Microextraction Coupled to Mass Spectrometry Via a Microfluidic Open Interface for Rapid Therapeutic Drug Monitoring

Overview
Journal Analyst
Specialty Chemistry
Date 2019 Apr 11
PMID 30968079
Citations 6
Authors
Affiliations
Soon will be listed here.
Abstract

Tranexamic acid (TXA) is an antifibrinolytic used during cardiac surgery that presents high inter-patient variability. High plasma concentrations have been associated with post-operative seizures. Due to the difficulties with maintaining acceptable concentrations of TXA during surgery, implementation of a point-of-care strategy for testing TXA plasma concentration would allow for close monitoring of its concentration during administration. This would facilitate timely corrections to the dosing schedule, and in effect tailor treatment for individual patient needs. In this work, a method for the rapid monitoring of TXA from plasma samples was subsequently carried out via biocompatible solid-phase microextraction (Bio-SPME) coupled directly to tandem mass spectrometry via a microfluidic open interface (MOI). MOI operates under the concept of a flow-isolated desorption volume and was designed with aims to directly hyphenate Bio-SPME to different detection and ionization systems. In addition, it allows the desorption of Bio-SPME fibers in small volumes while it concurrently continues feeding the ESI with a constant flow to minimize cross-talking and instabilities. The methodology was used to monitor six patients with varying degrees of renal dysfunction, at different time points during cardiac surgery. MOI proves to be a reliable and feasible tool for rapid therapeutic drug monitoring. Affording total times of analysis as low as 30 seconds per sample in its high throughput mode configuration while the single sample turn-around time was 15 minutes, including sample preparation. In addition, cross-validation against a standard thin film solid phase microextraction using liquid chromatography coupled to tandem mass spectrometry (TFME-LC-MS/MS) method was performed. Bland-Altman analysis was used to cross-validate the results obtained by the two methods. Data analysis demonstrated that 92% of the compared data pairs (n = 63) were distributed within the acceptable range. The data was also validated by the Passing Bablok regression, demonstrating good statistical agreement between these two methods. Finally, the currently presented method offers comparable results to the conventional liquid chromatography with acceptable RSDs, while only necessitating a fraction of the time. In this way, TXA concentration in plasma can be monitored in a close to real time throughput during surgery.

Citing Articles

Greenness assessment of microextraction techniques in therapeutic drug monitoring.

Pour P, Suzaei F, Daryanavard S Bioanalysis. 2024; 16(8):249-278.

PMID: 38466891 PMC: 11216521. DOI: 10.4155/bio-2023-0266.


solid phase microextraction for therapeutic monitoring and pharmacometabolomic fingerprinting of lung during lung perfusion of FOLFOX.

Looby N, Roszkowska A, Yu M, Rios-Gomez G, Pipkin M, Bojko B J Pharm Anal. 2023; 13(10):1195-1204.

PMID: 38024854 PMC: 10657970. DOI: 10.1016/j.jpha.2023.04.005.


Microfluidic nanodevices for drug sensing and screening applications.

Pal A, Kaswan K, Barman S, Lin Y, Chung J, Sharma M Biosens Bioelectron. 2022; 219:114783.

PMID: 36257116 PMC: 9533638. DOI: 10.1016/j.bios.2022.114783.


Solid-phase microextraction: a fit-for-purpose technique in biomedical analysis.

Bojko B Anal Bioanal Chem. 2022; 414(24):7005-7013.

PMID: 35606454 PMC: 9126758. DOI: 10.1007/s00216-022-04138-9.


Metabolomic Phenotyping of Gliomas: What Can We Get with Simplified Protocol for Intact Tissue Analysis?.

Gorynska P, Chmara K, Kupcewicz B, Gorynski K, Jaroch K, Paczkowski D Cancers (Basel). 2022; 14(2).

PMID: 35053475 PMC: 8773998. DOI: 10.3390/cancers14020312.