» Articles » PMID: 30948380

Image Quality of Low-Dose Cerebral Angiography and Effectiveness of Clinical Implementation on Diagnostic and Neurointerventional Procedures for Intracranial Aneurysms

Overview
Specialty Neurology
Date 2019 Apr 6
PMID 30948380
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Background And Purpose: Awareness of the potential for exposure to high doses of radiation from interventional radiologic procedures has increased. The purpose of this study was to evaluate image quality and dose reduction of low-dose cerebral angiography during diagnostic and therapeutic procedures for intracranial aneurysms.

Materials And Methods: A retrospective review of 1137 prospectively collected patients between January 2012 and June 2014 was performed. Beginning in April 2013, a dose-reduction strategy was implemented. Subjective image-quality assessment of 506 standard and 540 low-dose cerebral angiography images was performed by 2 neuroradiologists using a 5-point scale and was tested using noninferiority statistics. Radiation dose-area product and air kerma of 1046 diagnostic and 317 therapeutic procedures for intracranial aneurysms were analyzed and compared between groups before (group 1) and after (group 2) clinical implementation of a dose-reduction strategy.

Results: The image quality of the low-dose cerebral angiography was not inferior on the basis of results from the 2 readers. For diagnostic cerebral angiography, the mean dose-area product and air kerma were 140.8 Gy×cm and 1.0 Gy, respectively, in group 1 and 82.0 Gy×cm and 0.6 Gy in group 2 ( < .001, < .001). For the neurointerventional procedure, the mean dose-area product and air kerma were 246.0 Gy×cm and 3.7 Gy, respectively, in group 1 and 169.8 Gy×cm and 3.3 Gy in group 2 ( < .001, = .291).

Conclusions: With low-dose cerebral angiography, image quality was maintained, and implementation of dose-reduction strategies reduced radiation doses in patients undergoing diagnostic and neurointerventional procedures for intracranial aneurysms.

Citing Articles

Deep learning-based reconstruction for 3-dimensional heavily T2-weighted fat-saturated magnetic resonance (MR) myelography in epidural fluid detection: image quality and diagnostic performance.

Kim M, Yi J, Lee H, Hahn S, Lee Y, Lee J Quant Imaging Med Surg. 2024; 14(9):6531-6542.

PMID: 39281122 PMC: 11400679. DOI: 10.21037/qims-24-455.


Investigation of Fluoroscopy Time and Radiation Dose by the Number of Cerebral Angiography Operator Experiences.

Niimi J, Nakamura H, Ueda K, Yokoyama D, Tasaka K, Nemoto F J Neuroendovasc Ther. 2023; 16(4):187-191.

PMID: 37502451 PMC: 10370992. DOI: 10.5797/jnet.oa.2021-0043.


Low-Dose Three-Dimensional Rotational Angiography for Evaluating Intracranial Aneurysms: Analysis of Image Quality and Radiation Dose.

Ki H, Kim B, Kim J, Choi J, Shin Y, Choi Y Korean J Radiol. 2022; 23(2):256-263.

PMID: 35029071 PMC: 8814704. DOI: 10.3348/kjr.2021.0162.


Angular super-resolution in X-ray projection radiography using deep neural network: Implementation on rotational angiography.

Siow T, Ma C, Toh C Biomed J. 2022; 46(1):154-162.

PMID: 35026475 PMC: 10105049. DOI: 10.1016/j.bj.2022.01.001.


Low-Dose 3D Rotational Angiography in Measuring the Size of Intracranial Aneurysm: In Vitro Feasibility Study Using Aneurysm Phantom.

Ki H, Kim B, Kim J, Choi J, Shin Y, Choi Y Neurointervention. 2021; 16(1):59-63.

PMID: 33596630 PMC: 7946557. DOI: 10.5469/neuroint.2020.00437.

References
1.
Aroua A, Rickli H, Stauffer J, Schnyder P, Trueb P, Valley J . How to set up and apply reference levels in fluoroscopy at a national level. Eur Radiol. 2006; 17(6):1621-33. DOI: 10.1007/s00330-006-0463-3. View

2.
Pearl M, Torok C, Katz Z, Messina S, Blasco J, Tamargo R . Diagnostic quality and accuracy of low dose 3D-DSA protocols in the evaluation of intracranial aneurysms. J Neurointerv Surg. 2014; 7(5):386-90. DOI: 10.1136/neurintsurg-2014-011137. View

3.
DErcole L, Thyrion F, Bocchiola M, Mantovani L, Klersy C . Proposed local diagnostic reference levels in angiography and interventional neuroradiology and a preliminary analysis according to the complexity of the procedures. Phys Med. 2010; 28(1):61-70. DOI: 10.1016/j.ejmp.2010.10.008. View

4.
Ihn Y, Kim B, Byun J, Suh S, Won Y, Lee D . Patient Radiation Exposure During Diagnostic and Therapeutic Procedures for Intracranial Aneurysms: A Multicenter Study. Neurointervention. 2016; 11(2):78-85. PMC: 5018552. DOI: 10.5469/neuroint.2016.11.2.78. View

5.
Vano E, Fernandez J, Sanchez R, Martinez D, Ibor L, Gil A . Patient radiation dose management in the follow-up of potential skin injuries in neuroradiology. AJNR Am J Neuroradiol. 2012; 34(2):277-82. PMC: 7965086. DOI: 10.3174/ajnr.A3211. View