A Realistic Solution of Overcoming Disadvantage of Compton Suppression Spectrometer

Overview
Date 2019 Mar 20
PMID 30889529
Authors
Affiliations
Soon will be listed here.
Abstract

To achieve low detection limits for radionuclide analysis, Compton continuum suppression gamma-ray spectrometer has been used for several decades. The most serious disadvantage arises when radionuclides measured have cascade gamma emissions. This is a major limitation of the technique in environmental sample analysis. The digital list-mode data acquisition techniques in conjunction with time stamped pulse analysis method presented in this paper recovers the undesired full-energy-peak (FEP) coincidence losses for cascade radionuclides while retains suppression of true Compton events to keep the improved detection limit, thus extending the usefulness of the anti-Compton spectroscopy. In this study, software are developed to analyse the radionuclides with a relative simple decay scheme, such as Na, Co and Y, and a complex decay scheme, such as b, Cs, Eu and Eu. The results indicated that in the reconstruct anticoincidence spectra the FEP restoration efficiency, both for minor and major peaks, is near 100%.