Evaluating the Impact of Human Factors and Pen Needle Design on Insulin Pen Injection

Overview
Specialty Endocrinology
Date 2019 Mar 19
PMID 30880448
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Limited published data exists quantifying the influence of human factors (HF) and pen needle (PN) design on delivery outcomes of pen injection systems. This preclinical in vivo study examines the impact of PN hub design and applied force against the skin during injection on needle penetration depth (NPD).

Method: To precisely locate injection depth, PN injections (20 µl; 2 IU, U-100 volume equivalent) of iodinated contrast agent were administered to the flank of Yorkshire swine across a range of clinically relevant application forces against the skin (0.25, 0.75, 1.25, and 2.0 lbf). The NPD, representing in vivo needle tip depth in SC tissue, from four 32 G × 4 mm PN devices (BD Nano 2nd Gen and three commercial posted-hub PN devices; n = 75/device/force, 1200 total) was measured by fluoroscopic imaging of the resulting depot.

Results: The reengineered hub design more closely achieved the 4 mm target NPD with significantly less variability ( P = .006) than commercial posted-hub PN devices across the range of applied injection forces. Calculations of IM (intramuscular) injection risk completed through in silico probability model, using NPD and average human tissue thickness measurements, displayed a commensurate reduction (~2-8x) compared to conventional PN hub designs.

Conclusions: Quantifiable differences in injection depth were observed between identical labeled length PN devices indicating that hub design features, coupled with aspects of variable injection technique, may influence injection depth accuracy and consistency. The reengineered hub design may reduce the impact of unintended individual technique differences by improving target injection depth consistency and reducing IM injection potential.

Citing Articles

The Needle Shield Size and Applied Force of Subcutaneous Autoinjectors Significantly Influence the Injection Depth.

Staples A, Poulsen M, Praestmark K, Sparre T, Traberg M J Diabetes Sci Technol. 2024; :19322968241231996.

PMID: 38388411 PMC: 11571378. DOI: 10.1177/19322968241231996.


Optimising Insulin Injection Techniques to Improve Diabetes Outcomes.

Kalra S, Pathan F, Kshanti I, Bay N, Nagase T, Oliveria T Diabetes Ther. 2023; 14(11):1785-1799.

PMID: 37715887 PMC: 10570228. DOI: 10.1007/s13300-023-01460-y.


Imaging of large volume subcutaneous deposition using MRI: exploratory clinical study results.

Pettis R, Woodley W, Ossege K, Blum A, Bolick N, Rini C Drug Deliv Transl Res. 2023; 13(9):2353-2366.

PMID: 36913105 PMC: 10382358. DOI: 10.1007/s13346-023-01318-7.


Diabetes Technology Meeting 2021.

Xu N, Nguyen K, DuBord A, Pickup J, Sherr J, Teymourian H J Diabetes Sci Technol. 2022; 16(4):1016-1056.

PMID: 35499170 PMC: 9264449. DOI: 10.1177/19322968221090279.


Population Pharmacokinetics of Cotadutide in Subjects with Type 2 Diabetes.

Guan Y, Ly N, Li J, Arends R Clin Pharmacokinet. 2022; 61(6):833-845.

PMID: 35235191 DOI: 10.1007/s40262-021-01094-y.


References
1.
Vaag A, Pedersen K, Lauritzen M, Hildebrandt P, Beck-Nielsen H . Intramuscular versus subcutaneous injection of unmodified insulin: consequences for blood glucose control in patients with type 1 diabetes mellitus. Diabet Med. 1990; 7(4):335-42. DOI: 10.1111/j.1464-5491.1990.tb01401.x. View

2.
Norman J, Brown M, Raviele N, Prausnitz M, Felner E . Faster pharmacokinetics and increased patient acceptance of intradermal insulin delivery using a single hollow microneedle in children and adolescents with type 1 diabetes. Pediatr Diabetes. 2013; 14(6):459-65. DOI: 10.1111/pedi.12031. View

3.
Marcus A . Diabetes care - insulin delivery in a changing world. Medscape J Med. 2008; 10(5):120. PMC: 2438492. View

4.
Jiang S, Li P, Yu Y, Liu J, Yang Z . Experimental study of needle-tissue interaction forces: effect of needle geometries, insertion methods and tissue characteristics. J Biomech. 2014; 47(13):3344-53. DOI: 10.1016/j.jbiomech.2014.08.007. View

5.
Birkebaek N, Solvig J, Hansen B, Jorgensen C, Smedegaard J, Christiansen J . A 4-mm needle reduces the risk of intramuscular injections without increasing backflow to skin surface in lean diabetic children and adults. Diabetes Care. 2008; 31(9):e65. DOI: 10.2337/dc08-0977. View