Objective: To assess the technical performance (proportion of technical failure and unreliable measurements) of shear wave elastography (SWE) for assessing liver stiffness in pediatric and adolescent patients.

Methods: We searched Ovid-MEDLINE and EMBASE databases for eligible studies and selected original articles investigating transient elastography (TE), point shear wave elastography (pSWE), or two-dimensional SWE (2D-SWE) for measuring liver stiffness in pediatric and adolescent patients. A quantitative synthesis of studies reporting technical failures and/or unreliable measurements of TE, pSWE, or 2D-SWE is presented. Meta-analytic pooling was conducted using the random effects model. Meta-regression analysis was conducted to explore potential causes of heterogeneity.

Results: Forty of 69 studies (58%) provided technical performance information. Technical failure data were reported in 3 TE, 6 pSWE, and 8 2D-SWE studies. Unreliable measurement data were provided in 21 TE, 4 pSWE, and 1 2D-SWE study. The pooled proportion of unreliable measurements of TE was 12.1%. Meta-regression analysis showed that the study population size and readers' blinding to pathologic results affected the study's heterogeneity. The pooled proportions of technical failure during pSWE and 2D-SWE were 4.1% and 2.2%, respectively, demonstrating no significant difference between the techniques.

Conclusions: We reviewed the technical performance of SWE, especially the rate of unreliable measurements from TE studies and rates of technical failure from pSWE and 2D-SWE studies. Considering the importance of technical performance for clinical validation of SWE, numbers of and reasons for technical failure and unreliable measurements should be reported in future studies. Further efforts are necessary to standardize SWE reliability criteria.

Key Points: • Most TE studies reported rate of unreliable measurements, whereas pSWE and 2D-SWE studies were likely to report rates of technical failure. • The pooled proportion of unreliable measurements of TE was 12.1%. • The pooled proportions of technical failure during pSWE and 2D-SWE were 4.1% and 2.2%, respectively, demonstrating no significant difference between the techniques.

Citing Articles

The effect of orlistat in the treatment of non-alcoholic fatty liver in adolescents with overweight and obese.

Zahmatkesh A, Sohouli M, Shojaie S, Rohani P Eur J Pediatr. 2023; 183(3):1173-1182.

PMID: 38081992 DOI: 10.1007/s00431-023-05369-3.


Diagnostic assessment of two-dimensional shear wave elastography in relation to dimethyl arginine levels in dogs with chronic kidney disease.

Cho H, Yang S, Suh G, Choi J J Vet Sci. 2023; 24(6):e75.

PMID: 37904637 PMC: 10694370. DOI: 10.4142/jvs.23101.


Diagnostic Efficacy of Advanced Ultrasonography Imaging Techniques in Infants with Biliary Atresia (BA): A Systematic Review and Meta-Analysis.

Gunda S, Chambara N, Chen X, Pang M, Ying M Children (Basel). 2022; 9(11).

PMID: 36360404 PMC: 9688715. DOI: 10.3390/children9111676.


Hepatic Tumor Stiffness Measured by Shear Wave Elastography Is Prognostic for HCC Progression Following Treatment With Anti-PD-1 Antibodies Plus Lenvatinib: A Retrospective Analysis of Two Independent Cohorts.

Yuan G, Xie F, Song Y, Li Q, Li R, Hu X Front Immunol. 2022; 13:868809.

PMID: 35757765 PMC: 9218245. DOI: 10.3389/fimmu.2022.868809.


The Diagnostic Accuracy of LOGIQ S8 and E9 Shear Wave Elastography for Staging Hepatic Fibrosis, in Comparison with Transient Elastography.

Yoo J, Kim S, Kim Y Diagnostics (Basel). 2021; 11(10).

PMID: 34679515 PMC: 8535143. DOI: 10.3390/diagnostics11101817.


References
1.
Dietrich C, Bamber J, Berzigotti A, Bota S, Cantisani V, Castera L . EFSUMB Guidelines and Recommendations on the Clinical Use of Liver Ultrasound Elastography, Update 2017 (Long Version). Ultraschall Med. 2018; 38(4):e48. DOI: 10.1055/a-0641-0076. View

2.
Zhou L, Jiang H, Shan Q, Chen D, Lin X, Liu B . Liver stiffness measurements with supersonic shear wave elastography in the diagnosis of biliary atresia: a comparative study with grey-scale US. Eur Radiol. 2017; 27(8):3474-3484. DOI: 10.1007/s00330-016-4710-y. View

3.
Liberati A, Altman D, Tetzlaff J, Mulrow C, Gotzsche P, Ioannidis J . The PRISMA statement for reporting systematic reviews and meta-analyses of studies that evaluate health care interventions: explanation and elaboration. Ann Intern Med. 2009; 151(4):W65-94. DOI: 10.7326/0003-4819-151-4-200908180-00136. View

4.
Garcovich M, Veraldi S, Di Stasio E, Zocco M, Monti L, Toma P . Liver Stiffness in Pediatric Patients with Fatty Liver Disease: Diagnostic Accuracy and Reproducibility of Shear-Wave Elastography. Radiology. 2016; 283(3):820-827. DOI: 10.1148/radiol.2016161002. View

5.
Sagir A, Ney D, Oh J, Pandey S, Kircheis G, Mayatepek E . Evaluation of Acoustic Radiation Force Impulse Imaging (ARFI) for the Determination of Liver Stiffness Using Transient Elastography as a Reference in Children. Ultrasound Int Open. 2016; 1(1):E2-7. PMC: 5023198. DOI: 10.1055/s-0035-1554659. View