Accuracy of Fractional Flow Reserve Derived From Coronary Angiography

Abstract

Background: Measuring fractional flow reserve (FFR) with a pressure wire remains underutilized because of the invasiveness of guide wire placement or the need for a hyperemic stimulus. FFR derived from routine coronary angiography (FFR) eliminates both of these requirements and displays FFR values of the entire coronary tree. The FFR Accuracy versus Standard FFR (FAST-FFR) study is a prospective, multicenter, international trial with the primary goal of determining the accuracy of FFR.

Methods: Coronary angiography was performed in a routine fashion in patients with suspected coronary artery disease. FFR was measured in vessels with coronary lesions of varying severity using a coronary pressure wire and hyperemic stimulus. Based on angiograms of the respective arteries acquired in ≥2 different projections, on-site operators blinded to FFR then calculated FFR using proprietary software. Coprimary end points were the sensitivity and specificity of the dichotomously scored FFR for predicting pressure wire-derived FFR using a cutoff value of 0.80. The study was powered to meet prespecified performance goals for sensitivity and specificity.

Results: Ten centers in the United States, Europe, and Israel enrolled a total of 301 subjects and 319 vessels meeting inclusion/exclusion criteria which were included in the final analysis. The mean FFR was 0.81 and 43% of vessels had an FFR≤0.80. The per-vessel sensitivity and specificity were 94% (95% CI, 88% to 97%) and 91% (86% to 95%), respectively, both of which exceeded the prespecified performance goals. The diagnostic accuracy of FFR was 92% overall and remained high when only considering FFR values between 0.75 to 0.85 (87%). FFR values correlated well with FFR measurements ( r=0.80, P<0.001) and the Bland-Altman 95% confidence limits were between -0.14 and 0.12. The device success rate for FFR was 99%.

Conclusions: FFR measured from the coronary angiogram alone has a high sensitivity, specificity, and accuracy compared with pressure wire-derived FFR. FFR has the promise to substantially increase physiological coronary lesion assessment in the catheterization laboratory, thereby potentially leading to improved patient outcomes.

Clinical Trial Registration: URL: https://www.clinicaltrials.gov . Unique Identifier: NCT03226262.

Citing Articles

Non-invasive derivation of instantaneous free-wave ratio from invasive coronary angiography using a new deep learning artificial intelligence model and comparison with human operators' performance.

Oliveira C, Vilela M, Silva Marques J, Jorge C, Rodrigues T, Francisco A Int J Cardiovasc Imaging. 2025; .

PMID: 40063156 DOI: 10.1007/s10554-025-03369-y.


Enhancing quantitative coronary angiography (QCA) with advanced artificial intelligence: comparison with manual QCA and visual estimation.

Chae J, Kweon J, Park G, Park S, Yoon H, Lee C Int J Cardiovasc Imaging. 2025; 41(3):559-568.

PMID: 39875702 PMC: 11880186. DOI: 10.1007/s10554-025-03342-9.


Noninvasive machine-learning models for the detection of lesion-specific ischemia in patients with stable angina with intermediate stenosis severity on coronary CT angiography.

Hamasaki H, Arimura H, Yamasaki Y, Yamamoto T, Fukata M, Matoba T Phys Eng Sci Med. 2024; .

PMID: 39739189 DOI: 10.1007/s13246-024-01503-z.


Overview of Quantitative Flow Ratio and Optical Flow Ratio in the Assessment of Intermediate Coronary Lesions.

Westra J, Tu S US Cardiol. 2024; 14:e09.

PMID: 39720454 PMC: 11664755. DOI: 10.15420/usc.2020.09.


Cardiovascular Imaging in the Era of Precision Medicine: Insights from Advanced Technologies - A Narrative Review.

Manoel P, Dike I, Anis H, Yassin N, Wojtara M, Uwishema O Health Sci Rep. 2024; 7(11):e70173.

PMID: 39479287 PMC: 11522615. DOI: 10.1002/hsr2.70173.