Abstract

Background: The sensitivity of ELISA for biomarker detection can be significantly increased by integrating fluorescence with plasmonics. In surface-plasmon-coupled emission, the fluorophore emission is generally enhanced through the so-called physical mechanism due to an increase in the local electric field. Despite its fairly high enhancement factors, the use of surface-plasmon-coupled emission for high-throughput and point-of-care applications is still hampered due to the need for expensive focusing optics and spectrometers.

Methods: Here, we describe a new chemiplasmonic-sensing paradigm for enhanced emission through the molecular interactions between aromatic dyes and C films on Ag substrates.

Results: A 20-fold enhancement in the emission from rhodamine B-labeled biomolecules can be readily elicited without quenching its red color emission. As a proof of concept, we demonstrate two model bioassays using: 1) the RhB-streptavidin and biotin complexes in which the dye was excited using an inexpensive laser pointer and the ensuing enhanced emission was recorded by a smartphone camera without the need for focusing optics and 2) high-throughput 96-well plate assay for a model antigen (rabbit immunoglobulin) that showed detection sensitivity as low as 6.6 pM.

Conclusion: Our results show clear evidence that chemiplasmonic sensors can be extended to detect biomarkers in a point-of-care setting through a smartphone in simple normal incidence geometry without the need for focusing optics. Furthermore, chemiplasmonic sensors also facilitate high-throughput screening of biomarkers in the conventional 96-well plate format with 10-20 times higher sensitivity.

Citing Articles

Laboratory-on-a-Smartphone.

Mitsushio M Anal Sci. 2020; 36(2):141-142.

PMID: 32037345 DOI: 10.2116/analsci.highlights2002.

References
1.
Lakowicz J . Radiative decay engineering: biophysical and biomedical applications. Anal Biochem. 2001; 298(1):1-24. PMC: 6901132. DOI: 10.1006/abio.2001.5377. View

2.
Lakowicz J, Malicka J, Gryczynski I, Gryczynski Z . Directional surface plasmon-coupled emission: A new method for high sensitivity detection. Biochem Biophys Res Commun. 2003; 307(3):435-9. PMC: 2737392. DOI: 10.1016/s0006-291x(03)01214-2. View

3.
Lakowicz J . Radiative decay engineering 3. Surface plasmon-coupled directional emission. Anal Biochem. 2003; 324(2):153-69. PMC: 2737403. DOI: 10.1016/j.ab.2003.09.039. View

4.
Anger P, Bharadwaj P, Novotny L . Enhancement and quenching of single-molecule fluorescence. Phys Rev Lett. 2006; 96(11):113002. DOI: 10.1103/PhysRevLett.96.113002. View

5.
Fu Y, Lakowicz J . Enhanced fluorescence of Cy5-labeled oligonucleotides near silver island films: a distance effect study using single molecule spectroscopy. J Phys Chem B. 2006; 110(45):22557-62. PMC: 2737407. DOI: 10.1021/jp060402e. View