A New Device for Administration of Nasal Continuous Positive Airway Pressure in the Newborn: an Experimental Study

Overview
Journal Crit Care Med
Date 1988 Dec 1
PMID 3056654
Citations 20
Authors
Affiliations
Soon will be listed here.
Abstract

During treatment with continuous positive airway pressure (CPAP), optimal re-expansion of lung units with minimal work of breathing is best accomplished when the airway pressure (Paw) is kept constant at the desired CPAP level throughout the entire breathing cycle. To achieve this, a new device was constructed in which CPAP was generated by a jet of fresh gas close to the nasal airway. The performance of the new device was investigated experimentally using a lung model which simulated the breathing pattern of a newborn. Paw, flow, and external work of breathing were measured at three CPAP levels, with and without controlled airway leakage. The new device was compared with a traditional continuous-flow CPAP system with standard nasal prongs. Despite a virtually constant pressure within the traditional system, Paw variations and external workload were considerably less with the new device, which was also less sensitive to airway leakage.

Citing Articles

Non-invasive ventilation in neonates: a review of current literature.

Ramaswamy V, Devi R, Kumar G Front Pediatr. 2023; 11:1248836.

PMID: 38089684 PMC: 10713852. DOI: 10.3389/fped.2023.1248836.


Bubble devices versus other pressure sources for nasal continuous positive airway pressure in preterm infants.

Prakash R, De Paoli A, Davis P, Oddie S, McGuire W Cochrane Database Syst Rev. 2023; 3:CD015130.

PMID: 37009665 PMC: 10064833. DOI: 10.1002/14651858.CD015130.


Function of neonatal CPAP systems and the importance of total flow.

Jonsson B, Drevhammar T, Donaldsson S Pediatr Pulmonol. 2022; 58(1):18-19.

PMID: 36199238 PMC: 10092665. DOI: 10.1002/ppul.26195.


Breath-dependent pressure fluctuations in various constant- and variable-flow neonatal CPAP devices.

Auer-Hackenberg L, Stroicz P, Hofstatter E, Brandner J, Haselmann C, Wald M Pediatr Pulmonol. 2022; 57(10):2411-2419.

PMID: 35774021 PMC: 9796268. DOI: 10.1002/ppul.26050.


Comparison of inspiratory effort with three variable-flow nasal continuous positive airway pressure devices in preterm infants: a cross-over study.

Bordessoule A, Moreira A, Felice Civitillo C, Combescure C, Polito A, Rimensberger P Arch Dis Child Fetal Neonatal Ed. 2021; 106(4):404-407.

PMID: 33452219 PMC: 8237202. DOI: 10.1136/archdischild-2020-320531.