Translational Research on Clinically Failed Zirconia Implants

Overview
Journal Dent Mater
Specialty Dentistry
Date 2018 Dec 16
PMID 30551804
Citations 9
Authors
Affiliations
Soon will be listed here.
Abstract

Objectives: To provide fractographic analysis of clinically fractured zirconia implants recovered with their cemented crown. To calculate bending moments, corresponding stress and crack onset location on the implant's fracture surface using a mathematical model integrating spatial coordinates of the crown-implant part and occlusal loading obtained from 2D and 3D images.

Methods: 15 fractured zirconia implants parts (11 posterior and 4 anterior) with their all- ceramic crowns still cemented on it were recovered. The implants were first generations from four manufacturers (AXIS Biodental, Z-Systems, Straumann, Swiss Dental Solutions). The time-to-failure varied between 2weeks and 9years. Fractography was performed identifying the failure origin and characteristic surface crack features. From 2D and 3D digital images of the crown-implant part, spatial coordinates anchoring the crown's occlusal contacts with the implant's central axis and reference plane were integrated in a mathematical model spreadsheet. Loads of 500 N in total were selectively distributed over identified occlusal contacts from wear patterns. The resultant bending and torsion moments, corresponding shear, tensile, maximum principal stress and von Mises stress were calculated. The fracture crack onset location on the implant's fracture surface was given by an angular position with respect to an occlusal reference and compared with the location of the fracture origin identified from fractographic analysis.

Results: Implants fractured from the periphery of the smaller inner diameter between two threads at the bone-entrance level except for one implant which failed half-way within the bone. The porous coating (AXIS Biodental) and the large grit alumina sandblasting (Z-System) created surface defects directly related to the fracture origin. The model spreadsheet showed how occlusal loading with respect to the implant's central axis affects bending moments and crack onset. Dominant loads distributed on contacts with important wear pattern provided a calculated crack onset location in good agreement with the fractographic findings of the fracture origin.

Significance: Recovered broken zirconia implant parts with their restorative crowns can provide not only information regarding the failure origin using fractography but also knowledge regarding occlusal crown loading with respect to the implant's axis. The mathematical model was helpful in showing how occlusal loading affects the location of the fracture initiation site on clinical zirconia implant fracture cases.

Citing Articles

Investigation of Phase Transformation and Fracture Pattern as a Result of Long-Term Chewing Simulation and Static Loading of Reduced-Diameter Zirconia Implants.

Atalay Seckiner P, Gonuldas F, Akat B, Buyuksungur A, Orhan K Materials (Basel). 2024; 17(19).

PMID: 39410290 PMC: 11477572. DOI: 10.3390/ma17194719.


Zirconia Dental Implant Designs and Surface Modifications: A Narrative Review.

Ciszynski M, Chwaliszewski B, Simka W, Dominiak M, Gedrange T, Hadzik J Materials (Basel). 2024; 17(17).

PMID: 39274592 PMC: 11396535. DOI: 10.3390/ma17174202.


An epidemiological qualitative/quantitative SWOT-AHP analysis in order to highlight the positive or critical aspects of dental implants: A pilot study.

Aliberti S, Funk R, De Stefano M, Hoffmann T, Capunzo M Clin Exp Dent Res. 2024; 10(2):e2836.

PMID: 38450945 PMC: 10918715. DOI: 10.1002/cre2.836.


Clinical outcomes of zirconia implants: a systematic review and meta-analysis.

Mohseni P, Soufi A, Chrcanovic B Clin Oral Investig. 2023; 28(1):15.

PMID: 38135804 PMC: 10746607. DOI: 10.1007/s00784-023-05401-8.


Fatigue behavior of zirconia with microgrooved surfaces produced using femtosecond laser.

Li W, Ding Q, Sun F, Liu B, Yuan F, Zhang L Lasers Med Sci. 2023; 38(1):33.

PMID: 36598586 DOI: 10.1007/s10103-022-03679-w.