Effect of Screw Access Hole Design on the Fracture Resistance of Implant-supported Zirconia-based Restorations
Overview
Affiliations
Objective: To evaluate the effect of screw access hole design on the fracture resistance of cement-retained implant-supported zirconia-based restorations with screw access holes.
Materials And Methods: Thirty cement-retained implant-supported zirconia-based molar crown specimens were fabricated. The specimens were divided into 3 groups of 10: without access holes (C), with screw access holes (S), and with screw access holes accompanied with a surrounding zirconia wall (W). The veneering ceramic thickness was 0.8 mm in all groups. Implants were vertically mounted in a metal block. Abutments were screwed on the implants and the specimens were cemented to the abutments. A universal testing machine applied compressive forces to the specimens until fracture. Fracture resistance values of the specimens were measured. One-way analysis of variance (ANOVA) and Tukey HSD tests were used to analyze data (P < .05).
Results: The mean fracture resistance values were 5794.85 N for C, 2691.48 N for S, and 3878.06 N for W. The screw access hole design significantly affected the fracture resistance (P < .0001). Significant differences were found between C and S (P < .001), C and W (P < .001), and S and W (P = .026).
Conclusions: The screw access hole decreased the fracture resistance. A surrounding zirconia wall for the screw access hole increased the fracture resistance.
Clinical Significance: Screw access hole design may affect the fracture resistance of cement-retained implant-supported zirconia-based restorations with screw access holes. This study introduced a screw access hole design to improve the fracture resistance of these restorations.
Larsson A, Manuh J, Chrcanovic B Medicina (Kaunas). 2023; 59(9).
PMID: 37763722 PMC: 10535933. DOI: 10.3390/medicina59091603.
Zhou H, Ye S, Liu M, Feng H, Wen C BMC Oral Health. 2022; 22(1):108.
PMID: 35366875 PMC: 8976311. DOI: 10.1186/s12903-022-02145-x.
Wen C, Jiang R, Zhang Z, Lei B, Zhong Y, Zhou H Patient Prefer Adherence. 2022; 16:427-437.
PMID: 35210759 PMC: 8860758. DOI: 10.2147/PPA.S351143.
Farzin M, Derafshi R, Giti R, Taghva M J Dent (Shiraz). 2021; 22(2):132-137.
PMID: 34150950 PMC: 8206602. DOI: 10.30476/DENTJODS.2020.85080.1114.