Oscillometric and Doppler Arterial Blood Pressure Measurement in Conscious Goats

Overview
Journal Can J Vet Res
Date 2018 Oct 27
PMID 30363283
Citations 1
Authors
Affiliations
Soon will be listed here.
Abstract

The objective of this study was to characterize arterial blood pressure (BP) measurements obtained by using 2 indirect methods, oscillometry and Doppler ultrasonic sphygmomanometry, in conscious goats. Agreement between systolic BP yielded by these 2 methods was then assessed. Sixty female dairy goats aged from 1.5 to 11.8 y (median: 5.5 y) were examined in a standing position with a cuff placed on the tail. All goats had a severe arthritic form of caprine arthritis-encephalitis. Three to 5 repeated measurements of each BP type were averaged for each goat and considered as a final measurement. With oscillometry, systolic blood pressure (O-SBP), diastolic blood pressure, and mean blood pressure, as well as heart rate (HR) were measured, while only systolic blood pressure was measured with Doppler (D-SBP). The O-SBP did not correlate with D-SBP [correlation coefficient = 0.24, = 0.067]; the mean difference (± standard deviation) was 24.5 ± 26.3 mmHg and limits of agreement were from -27.2 mmHg [95% confidence interval (CI): -39.0, -15.4 mmHg] to 76.1 mmHg (95% CI: 64.3, 87.9 mmHg). No significant linear correlation was found between any BPs and HR ( -0.10 to 0.22) or age ( -0.26 to 0.07) of the goats. The study showed that, while BP could be measured in conscious goats using both oscillometry and Doppler ultrasonic sphygmomanometry, the results obtained were so inconsistent that these methods could not be used interchangeably.

Citing Articles

Comparison of oscillometric, Doppler and invasive blood pressure measurement in anesthetized goats.

Szalus-Jordanow O, Czopowicz M, Moroz A, Mickiewicz M, Garncarz M, Bagnicka E PLoS One. 2018; 13(5):e0197332.

PMID: 29791515 PMC: 5965870. DOI: 10.1371/journal.pone.0197332.

References
1.
Bailey J, Dunlop C, Chapman P, Demme W, Allen S, Heath R . Indirect Doppler ultrasonic measurement of arterial blood pressure results in a large measurement error in dorsally recumbent anaesthetised horses. Equine Vet J. 1994; 26(1):70-3. DOI: 10.1111/j.2042-3306.1994.tb04335.x. View

2.
Bodey A, Sansom J . Epidemiological study of blood pressure in domestic cats. J Small Anim Pract. 1999; 39(12):567-73. DOI: 10.1111/j.1748-5827.1998.tb03710.x. View

3.
da Cunha A, Saile K, Beaufrere H, Wolfson W, Seaton D, Acierno M . Measuring level of agreement between values obtained by directly measured blood pressure and ultrasonic Doppler flow detector in cats. J Vet Emerg Crit Care (San Antonio). 2014; 24(3):272-8. DOI: 10.1111/vec.12161. View

4.
McMurphy R, Stoll M, McCubrey R . Accuracy of an oscillometric blood pressure monitor during phenylephrine-induced hypertension in dogs. Am J Vet Res. 2006; 67(9):1541-5. DOI: 10.2460/ajvr.67.9.1541. View

5.
Pedersen K, Butler M, Ersboll A, Pedersen H . Evaluation of an oscillometric blood pressure monitor for use in anesthetized cats. J Am Vet Med Assoc. 2002; 221(5):646-50. DOI: 10.2460/javma.2002.221.646. View