Transcoronary Pacing : Reliability During Myocardial Ischemia and After Implantation of a Coronary Stent

Overview
Date 2018 Oct 11
PMID 30302524
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Transcoronary pacing is a useful therapeutic option for the treatment of unheralded bradycardias in the setting of percutaneous coronary interventions (PCI).

Objectives: In the present study, we investigated the influence of stent implantation and transient myocardial ischemia on the feasibility of transcoronary pacing in a porcine model.

Methods: 7 adult pigs underwent a percutaneous coronary intervention with implantation of a coronary stent under general anaesthesia in an animal catheterization laboratory. Transcoronary pacing was established by using a standard guidewire isolated with an angioplasty balloon positioned in the periphery of the right coronary artery serving as the cathode. As the indifferent anode, a skin patch electrode at the back of the animal was used. The reliability of transcoronary pacing was assessed by measurement of threshold and impedance data and the magnitude of the epicardial electrogram at baseline, after implantation of a coronary stent and finally during myocardial ischemia.

Results: Effective transcoronary pacing could be demonstrated in all cases with the standard unipolar transcoronary pacing setup yielding a low pacing threshold at baseline of 1.3 ± 0.8 V with an impedance of 283 ± 67 Ω. Implantation of a coronary stent did not influence the pacing threshold (1.0 ± 0.4 V) and impedance (262 ± 63 Ω). Acute myocardial ischemia lead to a significant but clinically nonrelevant increase of the pacing threshold to 2.0 ± 0.6 V and a drop in pacing impedance (137 ± 39 Ω).

Conclusions: Transcoronary pacing in the animal model is not affected by implantation of a coronary stent in the same vessel used for pacing. Despite a significant increase in pacing threshold, the transcoronary pacing approach is reliable in acute myocardial ischemia during a percutaneous coronary intervention.

Citing Articles

Effects of coronary artery disease in patients with permanent left bundle branch area pacing: A retrospective study.

Shan Y, Lin M, Ye M, Shen X, Li D, Chen Z Heliyon. 2024; 10(2):e24226.

PMID: 38268827 PMC: 10803899. DOI: 10.1016/j.heliyon.2024.e24226.


Coronary Guidewires in Temporary Cardiac Pacing and Assessment of Myocardial Viability: Current Perspectives and Future Directions.

Javid R, Wassef N, Wheatcroft S, Tayebjee M J Clin Med. 2023; 12(22).

PMID: 38002591 PMC: 10672001. DOI: 10.3390/jcm12226976.


Intracoronary Pacing during "Chimney Technique" in Transcatheter Aortic Valve-in-Valve Implantation: An Alternative Temporary Rapid Ventricular Stimulation?.

Cafaro A, Rizzo F, Fischetti D, Quarta L, Mussardo M, Mandurino-Mirizzi A J Cardiovasc Dev Dis. 2023; 10(8).

PMID: 37623354 PMC: 10455771. DOI: 10.3390/jcdd10080341.

References
1.
Meier B, RUTISHAUSER W . Coronary pacing during percutaneous transluminal coronary angioplasty. Circulation. 1985; 71(3):557-61. DOI: 10.1161/01.cir.71.3.557. View

2.
Chatelain P, Meier B, Belenger J, Killisch J, Cox J, RUTISHAUSER W . [Emergency cardiac pacing via a coronary vessel during percutaneous coronary angioplasty]. Arch Mal Coeur Vaiss. 1985; 78(10):1583-7. View

3.
Asano M, Mishima A, Ishii T, Takeuchi Y, Suzuki Y, Manabe T . Surgical treatment for right ventricular perforation caused by transvenous pacing electrodes: a report of three cases. Surg Today. 1996; 26(11):933-5. DOI: 10.1007/BF00311800. View

4.
Laird J, Hull R, Stajduhar K, Weston L, Kufs W, Wortham D . Transcoronary cardiac pacing during myocardial ischemia. Cathet Cardiovasc Diagn. 1993; 30(2):162-5. DOI: 10.1002/ccd.1810300217. View

5.
Ohm O, Mitamura H, Michelson E, Sauermelch C, DREIFUS L . Ventricular tachyarrhythmia initiation in a canine model of recent myocardial infarction. Comparison of unipolar cathodal, anodal and bipolar stimulation. Cardiology. 1987; 74(3):169-81. DOI: 10.1159/000174195. View