Accuracy and Precision of Transcutaneous Carbon Dioxide Monitoring: a Systematic Review and Meta-analysis

Overview
Journal Thorax
Date 2018 Sep 14
PMID 30209079
Citations 18
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Transcutaneous carbon dioxide (TcCO) monitoring is a non-invasive alternative to arterial blood sampling. The aim of this review was to determine the accuracy and precision of TcCO measurements.

Methods: Medline and EMBASE (2000-2016) were searched for studies that reported on a measurement of PaCO that coincided with a measurement of TcCO. Study selection and quality assessment (using the revised Quality Assessment of Diagnostic Accuracy Studies tool (QUADAS-2)) were performed independently. The Grading Quality of Evidence and Strength of Recommendation approach was used to summarise the strength of the body of evidence. Pooled estimates of the mean bias between TcCO and PaCO and limits of agreement with outer 95% CIs (termed population limits of agreement) were calculated.

Results: The mean bias was -0.1 mm Hg and the population limits of agreement were -15 to 15 mm Hg for 7021 paired measurements taken from 2817 participants in 73 studies, which was outside of the clinically acceptable range (7.5 mm Hg). The lowest PaCO reported in the studies was 18 mm Hg and the highest was 103 mm Hg. The major sources of inconsistency were sensor location and temperature. The population limits of agreement were within the clinically acceptable range across 3974 paired measurements from 1786 participants in 44 studies that applied the sensor to the earlobe using the TOSCA and Sentec devices (-6 to 6 mm Hg).

Conclusion: There are substantial differences between TcCO and PaCO depending on the context in which this technology is used. TcCO sensors should preferentially be applied to the earlobe and users should consider setting the temperature of the sensor higher than 42°C when monitoring at other sites.

Systematic Review Registration Number: PROSPERO; CRD42017057450.

Citing Articles

Efficacy of Noninvasive Ventilation With Expiratory Washout in Stable COPD Patients.

Kung S, Semprini A, Kirton L, Fogarin J, Zoellner S, Beasley R Respir Care. 2025; 70(1):29-37.

PMID: 39964871 PMC: 11824877. DOI: 10.1089/respcare.11876.


Identifying Daytime Hypercapnia Using Transcutaneous Carbon Dioxide Monitoring in Patients With Amyotrophic Lateral Sclerosis.

Morrison A, Jimenez J, Hsu J, Elman L, Choi P, Ackrivo J Muscle Nerve. 2025; 71(4):611-619.

PMID: 39936179 PMC: 11887530. DOI: 10.1002/mus.28366.


Accuracy and Interpretation of Transcutaneous Carbon Dioxide Monitoring in Critically Ill Children.

Setar L, Lee J, Sanchez-Pinto L, Coates B Pediatr Crit Care Med. 2024; 25(9):e372-e379.

PMID: 39288436 PMC: 11368163. DOI: 10.1097/PCC.0000000000003564.


Update in Noninvasive Home Mechanical Ventilation: A Narrative Review of Indications, Outcomes, and Monitoring.

Tregidgo L, Naran P, Gosal E, DCruz R Can Respir J. 2024; 2024:7013576.

PMID: 38989047 PMC: 11236466. DOI: 10.1155/2024/7013576.


Evaluation of agreement between a noninvasive method for real-time measurement of critical blood values with a standard point-of-care device.

Ettrich R, Caballero J, Sakharkar P, Ahmed S, Hurlston T, Parmar J PLoS One. 2024; 19(6):e0304706.

PMID: 38889168 PMC: 11185484. DOI: 10.1371/journal.pone.0304706.