Comparative Evaluation of Accuracy of Two Electronic Apex Locators in the Presence of Contemporary Irrigants: An Study

Overview
Date 2018 Aug 21
PMID 30123769
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Aim: The present study was aimed to compare the accuracy of Root ZX Mini and Propex II in the presence of 0.1% octinidine dohydrochloride (OCT), 2% chlorhexidine gluconate (CHX), and 5% sodium hypochlorite (NaOCl) heated and nonheated before and after preparation.

Materials And Methods: Eighty extracted single-rooted teeth were selected for the study and decoronated. Teeth were mounted in an alginate model. Actual working length (AL) was measured using a stereomicroscope under ×4 magnification. Electronic working length measurements were recorded using Root ZX Mini and Propex II apex locators in the presence of 0.1% OCT, 2% CHX, and 5% NaOCl (nonheated and heated to 60°C) before and after preparation. Mean and standard deviation differences before and after preparation were calculated and statistically analyzed using analysis of variance and paired -test.

Results: The accuracy of Root ZX Mini before and after preparation within ±0.5 mm of AL was consistently high in the presence of irrigants than Propex II. 5% NaOCl (heated and nonheated) showed more variation than the other irrigants, in the working length determination in both the apex locators.

Conclusion: Electronic length measurements were shorter with heated and nonheated 5% NaOCl and longer with 0.1% OCT and 2% CHX for both the electronic apex locators.

Citing Articles

The effect of different root canal irrigations on the accuracy of apex locators: A systematic review.

Shekarbaghani S, Bolhari B, Khalilak Z J Clin Exp Dent. 2025; 16(12):e1538-e1546.

PMID: 39822780 PMC: 11733891. DOI: 10.4317/jced.62049.


Effect of Different Electroconductive Root Canal Irrigations on the Accuracy of Different Apex Locators: An Comparative Study.

Yahya E, Alchalabi A, Alkhalidi E J Int Soc Prev Community Dent. 2024; 14(3):211-217.

PMID: 39055293 PMC: 11268530. DOI: 10.4103/jispcd.jispcd_192_23.


Accuracy of an Electronic Apex Locator (EAL) in Measuring the Working Length of Root Canals in Comparison With Radiographs: An In Vivo Study.

Bhagat E, Agnihotri Y, Das A, Das S, Panda S, Hota S Cureus. 2024; 16(5):e60283.

PMID: 38872664 PMC: 11173343. DOI: 10.7759/cureus.60283.


Effect of Sodium Hypochlorite Concentration on Electronic Apex Locator Reliability.

Diemer F, Plews E, Georgelin-Gurgel M, Mishra L, Kim H Materials (Basel). 2022; 15(3).

PMID: 35160807 PMC: 8836685. DOI: 10.3390/ma15030863.


Comparison of Accuracy in Determining the Root Canal Working Length by Using Two Generations of Apex Locators - An In Vitro Study.

Golvankar K, Kader M, Latheef A, Ali A, Abullais S, Sibagathullah M Open Access Maced J Med Sci. 2020; 7(19):3276-3280.

PMID: 31949530 PMC: 6953924. DOI: 10.3889/oamjms.2019.696.


References
1.
Jain S, Kapur R . Comparative evaluation of accuracy of two electronic apex locators in the presence of various irrigants: An in vitro study. Contemp Clin Dent. 2012; 3(Suppl 2):S140-5. PMC: 3514940. DOI: 10.4103/0976-237X.101068. View

2.
Khattak O, Raidullah E, Francis M . A comparative assessment of the accuracy of electronic apex locator (Root ZX) in the presence of commonly used irrigating solutions. J Clin Exp Dent. 2014; 6(1):e41-6. PMC: 3935904. DOI: 10.4317/jced.51230. View

3.
Duran-Sindreu F, Gomes S, Stober E, Mercade M, Jane L, Roig M . In vivo evaluation of the iPex and Root ZX electronic apex locators using various irrigants. Int Endod J. 2013; 46(8):769-74. DOI: 10.1111/iej.12057. View

4.
Sirtes G, Waltimo T, Schaetzle M, Zehnder M . The effects of temperature on sodium hypochlorite short-term stability, pulp dissolution capacity, and antimicrobial efficacy. J Endod. 2005; 31(9):669-71. DOI: 10.1097/01.don.0000153846.62144.d2. View

5.
Ozsezer E, Inan U, Aydin U . In vivo evaluation of ProPex electronic apex locator. J Endod. 2007; 33(8):974-7. DOI: 10.1016/j.joen.2007.04.003. View