» Articles » PMID: 30106900

Observation of Short Wavelength Infrared (SWIR) Cherenkov Emission

Overview
Journal Opt Lett
Specialty Ophthalmology
Date 2018 Aug 15
PMID 30106900
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Cherenkov emission induced by external beam radiation from a clinical linear accelerator has been shown in preclinical molecular imaging and clinical imaging. The broad spectrum Cherenkov emission should have a short wavelength infrared (SWIR, 1000-1700 nm) component, as predicted theoretically. To the best of our knowledge, this Letter is the first experimental observation of this SWIR Cherenkov emission induced by external beam radiation. The measured spectrum of SWIR Cherenkov emission matches the theoretical prediction, with a fluence rate near one-third of the visible and near-infrared red emissions (Vis-NIR, 400-900 nm). Imaging in water-based phantoms and biological tissues indicates that there is a sufficient fluence rate for radiotherapy dosimetry applications. The spatial resolution is improved approximately 5.3 times with SWIR Cherenkov emission detection versus Vis-NIR Cherenkov emission, which provides some improvement in the potential for higher resolution Cherenkov emission dosimetry and molecular sensing during clinical radiotherapy by imaging with SWIR wavelengths.

Citing Articles

A review of recent and emerging approaches for the clinical application of Cerenkov luminescence imaging.

Larney B, Skubal M, Grimm J Front Phys. 2023; 9.

PMID: 36845872 PMC: 9957555. DOI: 10.3389/fphy.2021.684196.


Detection of Shortwave-Infrared Cerenkov Luminescence from Medical Isotopes.

Mc Larney B, Zhang Q, Pratt E, Skubal M, Isaac E, Hsu H J Nucl Med. 2022; 64(1):177-182.

PMID: 35738902 PMC: 9841262. DOI: 10.2967/jnumed.122.264079.


Remote dose imaging from Cherenkov light using spatially resolved CT calibration in breast radiotherapy.

Hachadorian R, Bruza P, Jermyn M, Gladstone D, Zhang R, Jarvis L Med Phys. 2022; 49(6):4018-4025.

PMID: 35304768 PMC: 9187603. DOI: 10.1002/mp.15614.


Review of in vivo optical molecular imaging and sensing from x-ray excitation.

Pogue B, Zhang R, Cao X, Jia J, Petusseau A, Bruza P J Biomed Opt. 2021; 26(1).

PMID: 33386709 PMC: 7778455. DOI: 10.1117/1.JBO.26.1.010902.


Correcting Cherenkov light attenuation in tissue using spatial frequency domain imaging for quantitative surface dosimetry during whole breast radiation therapy.

Hachadorian R, Bruza P, Jermyn M, Mazhar A, Cuccia D, Jarvis L J Biomed Opt. 2018; 24(7):1-10.

PMID: 30415511 PMC: 6228320. DOI: 10.1117/1.JBO.24.7.071609.

References
1.
Glaser A, Zhang R, Davis S, Gladstone D, Pogue B . Time-gated Cherenkov emission spectroscopy from linear accelerator irradiation of tissue phantoms. Opt Lett. 2012; 37(7):1193-5. PMC: 3698962. DOI: 10.1364/OL.37.001193. View

2.
Hong G, Zou Y, Antaris A, Diao S, Wu D, Cheng K . Ultrafast fluorescence imaging in vivo with conjugated polymer fluorophores in the second near-infrared window. Nat Commun. 2014; 5:4206. DOI: 10.1038/ncomms5206. View

3.
Pogue B, Feng J, LaRochelle E, Bruza P, Lin H, Zhang R . Maps of in vivo oxygen pressure with submillimetre resolution and nanomolar sensitivity enabled by Cherenkov-excited luminescence scanned imaging. Nat Biomed Eng. 2019; 2(4):254-264. PMC: 6424530. DOI: 10.1038/s41551-018-0220-3. View

4.
Andreozzi J, Zhang R, Gladstone D, Williams B, Glaser A, Pogue B . Cherenkov imaging method for rapid optimization of clinical treatment geometry in total skin electron beam therapy. Med Phys. 2016; 43(2):993-1002. PMC: 4744235. DOI: 10.1118/1.4939880. View

5.
Welsher K, Sherlock S, Dai H . Deep-tissue anatomical imaging of mice using carbon nanotube fluorophores in the second near-infrared window. Proc Natl Acad Sci U S A. 2011; 108(22):8943-8. PMC: 3107273. DOI: 10.1073/pnas.1014501108. View