Purpose: In this study, lung radiotherapy target volumes as well as critical organs such as the lungs, spinal cord, esophagus, and heart doses calculated using pencil beam (PB) and Monte Carlo (MC) algorithm-based treatment planning systems (TPSs) were compared. The main aim was the evaluation of calculated dose differences between the PB and MC algorithms in a highly heterogeneous medium.

Methods: A total of 6 MV photon energy conformal treatment plans were created for a RANDO lung phantom using one PB algorithm-based Precise Plan Release 2.16 TPS and one MC algorithm-based Monaco TPS. Thermoluminescence dosimeters (TLDs) were placed into appropriate slices within the RANDO phantom and then irradiated with an Elekta-Synergy Linear Accelerator for dose verification. Doses were calculated for the V5, V10, V20, and mean lung doses (MLDs) in bilateral lungs and D50, D98, D2, and mean doses in the target volume (planning target volume, PTV).

Results: The minimum, maximum, and mean doses of the target volumes and critical organs in two treatment plans were compared using dose volume histograms (DVHs). The mean dose difference between the PB and MC algorithms for the PTV was 0.3%, whereas the differences in V5, V10, V20, and MLD were 12.5%, 15.8%, 14.4%, and 9.1%, respectively. The differences in PTV coverage between the two algorithms were 0.9%, 2.7% and 0.7% for D50, D98 and D2, respectively.

Conclusions: A comparison of the dose data acquired in this study reveals that the MC algorithm calculations are closer to the 60 Gy prescribed dose for PTV, while the difference between the PB and MC algorithms was found to be non-significant. Because of the major difference arising from the dose calculation techniques by TPS that was observed in the MLD with significant medium heterogeneity, we recommend the use of the MC algorithm in such heterogeneous sites.

Citing Articles

Monte Carlo methods to assess biological response to radiation in peripheral organs and in critical organs near the target.

Matuszak N, Piotrowski I, Kruszyna-Mochalska M, Skrobala A, Mocydlarz-Adamcewicz M, Malicki J Rep Pract Oncol Radiother. 2025; 29(5):638-648.

PMID: 39759550 PMC: 11698553. DOI: 10.5603/rpor.103525.


Dosimetric Impact of Prescription Point Placement in Heterogeneous Medium for Conformal Radiotherapy Dose Calculation with Various Algorithms.

Pandu B, Khanna D, Palanisamy M, Jacob S, Manichan S J Med Phys. 2024; 49(3):400-409.

PMID: 39526146 PMC: 11548077. DOI: 10.4103/jmp.jmp_71_24.


Impact of Dose Calculation Algorithms and Radiobiological Parameters on Prediction of Cardiopulmonary Complications in Left Breast Radiation Therapy.

Kargar N, Zeinali A, Molazadeh M J Biomed Phys Eng. 2024; 14(2):129-140.

PMID: 38628897 PMC: 11016826. DOI: 10.31661/jbpe.v0i0.2305-1616.


Dosimetric accuracy of three dose calculation algorithms for radiation therapy of in situ non-small cell lung carcinoma.

Kolacio M, Rajlic D, Radojcic M, Radojcic D, Obajdin N, Dundara Debeljuh D Rep Pract Oncol Radiother. 2022; 27(1):86-96.

PMID: 35402037 PMC: 8989458. DOI: 10.5603/RPOR.a2022.0013.


Investigation of tube voltage dependence on CT number and its effect on dose calculation algorithms using thorax phantom in Monaco treatment planning system for external beam radiation therapy.

Saini A, Pandey V, Kumar P, Singh A, Pasricha R J Med Phys. 2022; 46(4):315-323.

PMID: 35261502 PMC: 8853450. DOI: 10.4103/jmp.JMP_124_20.


References
1.
Nielsen T, Wieslander E, Fogliata A, Nielsen M, Hansen O, Brink C . Influence of dose calculation algorithms on the predicted dose distribution and NTCP values for NSCLC patients. Med Phys. 2011; 38(5):2412-8. DOI: 10.1118/1.3575418. View

2.
Andreo P . Monte Carlo techniques in medical radiation physics. Phys Med Biol. 1991; 36(7):861-920. DOI: 10.1088/0031-9155/36/7/001. View

3.
Bufacchi A, Nardiello B, Capparella R, Begnozzi L . Clinical implications in the use of the PBC algorithm versus the AAA by comparison of different NTCP models/parameters. Radiat Oncol. 2013; 8(1):164. PMC: 3750611. DOI: 10.1186/1748-717X-8-164. View

4.
Rogers D . Fifty years of Monte Carlo simulations for medical physics. Phys Med Biol. 2006; 51(13):R287-301. DOI: 10.1088/0031-9155/51/13/R17. View

5.
Ali I, Ahmad S . Quantitative assessment of the accuracy of dose calculation using pencil beam and Monte Carlo algorithms and requirements for clinical quality assurance. Med Dosim. 2013; 38(3):255-61. DOI: 10.1016/j.meddos.2013.02.005. View