Safe Pseudoelastic Limit Range Under Torsional Loading with Reciproc Blue

Overview
Journal Int Endod J
Specialty Dentistry
Date 2018 Jul 25
PMID 30040121
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

Aim: To determine the distortional angle and torsional load at the pseudoelastic limit of Reciproc Blue instruments and to verify the safety of using pre-set dedicated motors designed for use with the original Reciproc instruments.

Methodology: Two torsional conditions of Reciproc R25 and Reciproc Blue R25 were tested using a custom device. The first condition fixed the file tips at 3 mm and repetitively rotated them, with gradually increasing angles, from 10° to 270°, and the resulting torque was recorded (n = 15). The second test involved a single continuous rotation until fracture (n = 15). The pseudoelastic limits of the instruments were determined from their torque-rotation curves. For statistical analysis, two-way analysis of variance and t-tests were used, at a 95% significance level. Tested specimens were examined using field-emission scanning electron microscopy (FE-SEM).

Results: The angle at the pseudoelastic limit was significantly greater for Reciproc Blue than Reciproc, regardless of test mode (P < 0.05). When torsional loading was repeated, using gradually increasing rotational angles, the torsional resistance was significantly lower than for a single rotation (P < 0.05). However, under all test conditions, the pseudoelastic limit was below the pre-set 170° of the dedicated reciprocating motor. FE-SEM evaluation of the lateral aspects of the instruments revealed numerous longitudinal microcracks running along their long axis. For Reciproc Blue groups, the file shaft machining grooves were distorted after repetitive and continuous torsional tests.

Conclusions: The 170° pre-set angle of the dedicated endodontic motors for the Reciproc system was safe for Reciproc Blue in single or time-restricted use.

Citing Articles

Shaping outcomes of Reciproc Blue and Rotate in roots canals with band-shaped isthmuses: micro-CT study.

Keskin C, Pirimoglu B, Ciftcioglu E, Dinger E, Komec O, Keles A Clin Oral Investig. 2023; 27(12):7337-7344.

PMID: 37861748 DOI: 10.1007/s00784-023-05324-4.


Torsional Resistance of Heat-Treated Nickel-Titanium Instruments under Different Temperature Conditions.

Jo H, Kwak S, Kim H, Kim S, Ha J Materials (Basel). 2021; 14(18).

PMID: 34576520 PMC: 8470266. DOI: 10.3390/ma14185295.


Influence of the type of reciprocating motion on the cyclic fatigue resistance of reciprocating files in a dynamic model.

Zubizarreta-Macho A, Albaladejo Martinez A, Falcao Costa C, Quispe-Lopez N, Agustin-Panadero R, Mena-Alvarez J BMC Oral Health. 2021; 21(1):179.

PMID: 33827530 PMC: 8028824. DOI: 10.1186/s12903-021-01538-8.