Optimization of Parameters for Femoral Component Implantation During TKA Using Finite Element Analysis and Orthogonal Array Testing

Overview
Publisher Biomed Central
Specialty Orthopedics
Date 2018 Jul 22
PMID 30029670
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Individualized and accurate implantation of a femoral component during total knee arthroplasty (TKA) is essential in achieving equal distribution of intra-articular stress and long-term survival of the prosthesis. However, individualized component implantation remains challenging. This study aimed to optimize and individualize the positioning parameters of a femoral component in order to facilitate its accurate implantation.

Methods: Using computer-simulated TKA, the positioning parameters of a femoral component were optimized individually by finite element analysis in combination with orthogonal array testing. Flexion angle, valgus angle, and external rotation angle were optimized in order to reduce the peak value of the pressure on the polyethylene liner of the prosthesis.

Results: The optimal implantation parameters of the femoral component were as follows: 1° flexion, 5° valgus angle, and 4° external rotation. Under these conditions, the peak value of the pressure on the polyethylene liner surface was minimized to 16.46 MPa. Among the three parameters, the external rotation angle had the greatest effect on the pressure, followed by the valgus angle and the flexion angle.

Conclusion: Finite element analysis in combination with orthogonal array testing can optimize the implantation parameters of a femoral component for TKA. This approach would possibly reduce the wear of the polyethylene liner and prolong the survival of the TKA prosthesis, due to its capacity to minimize stress. This technique represents a new method for preoperative optimization of the implantation parameters that can achieve the best possible TKA outcome.

Citing Articles

Finite element analysis and its application in Orthopaedics: A narrative review.

Verma A, Jain A, Sethy S, Verma V, Goyal N, Vathulya M J Clin Orthop Trauma. 2025; 58():102803.

PMID: 39781401 PMC: 11703707. DOI: 10.1016/j.jcot.2024.102803.


Total Knee Replacement: Subject-Specific Modeling, Finite Element Analysis, and Evaluation of Dynamic Activities.

Loi I, Stanev D, Moustakas K Front Bioeng Biotechnol. 2021; 9:648356.

PMID: 33937216 PMC: 8085535. DOI: 10.3389/fbioe.2021.648356.


Comparison of the contact stress between the sensor and real polyethylene insert in total knee arthroplasty: a finite element analysis.

Song S, Kim K, Park C Ann Transl Med. 2020; 8(21):1424.

PMID: 33313169 PMC: 7723577. DOI: 10.21037/atm-20-3011.


The effects of immediate programmed cryotherapy and continuous passive motion in patients after computer-assisted total knee arthroplasty: a prospective, randomized controlled trial.

Chen M, Lin C, Ko J, Kuo F J Orthop Surg Res. 2020; 15(1):379.

PMID: 32883309 PMC: 7469108. DOI: 10.1186/s13018-020-01924-y.

References
1.
Ji H, Han J, Jin D, Seo H, Won Y . Kinematically aligned TKA can align knee joint line to horizontal. Knee Surg Sports Traumatol Arthrosc. 2016; 24(8):2436-41. DOI: 10.1007/s00167-016-3995-3. View

2.
Weiss J, Gardiner J . Computational modeling of ligament mechanics. Crit Rev Biomed Eng. 2001; 29(3):303-71. DOI: 10.1615/critrevbiomedeng.v29.i3.20. View

3.
Li Z, Esposito C, Koch C, Lee Y, Padgett D, Wright T . Polyethylene Damage Increases With Varus Implant Alignment in Posterior-stabilized and Constrained Condylar Knee Arthroplasty. Clin Orthop Relat Res. 2017; 475(12):2981-2991. PMC: 5670063. DOI: 10.1007/s11999-017-5477-4. View

4.
Allen C, Hooper G, Oram B, Wells J . Does computer-assisted total knee arthroplasty improve the overall component position and patient function?. Int Orthop. 2013; 38(2):251-7. PMC: 3923953. DOI: 10.1007/s00264-013-1996-0. View

5.
Liau J, Cheng C, Huang C, Lo W . The effect of malalignment on stresses in polyethylene component of total knee prostheses--a finite element analysis. Clin Biomech (Bristol). 2002; 17(2):140-6. DOI: 10.1016/s0268-0033(01)00109-7. View