Validation of Optimum ROI Size for I-FP-CIT SPECT Imaging Using a 3D Mathematical Cylinder Phantom

Overview
Specialty Nuclear Medicine
Date 2018 Jul 13
PMID 29998147
Authors
Affiliations
Soon will be listed here.
Abstract

Objectives: The partial volume effect (PVE) of single-photon emission computed tomography (SPECT) on corpus striatum imaging is caused by the underestimation of specific binding ratio (SBR). A large ROI (region of interest) set using the Southampton method is independent of PVE for SBR. The present study aimed to determine the optimal ROI size with contrast and SBR for striatum images and validate the Southampton method using a three-dimensional mathematical cylinder (3D-MAC) phantom.

Methods: We used ROIs sizes of 27, 36, 44, 51, 61, 68, and 76 mm for targets with diameters 40, 20, and 10 mm on reference and processed images reconstructed using the 3D-MAC phantom. Contrast values and SBR were compared with the theoretical values to obtain the optimal ROI size.

Results: The contrast values in the ROI with diameters of 51 (target: 40 mm in diameter) and 44 (target: 20 mm in diameter) mm matched the theoretical values. However, this value did not correspond with the 10-mm-diameter target. The SBR matched the theoretical value with an ROI of > 44 mm in the 20-mm-diameter target; but, it was under- and overestimated under any other conditions.

Conclusion: These results suggested that an ROI should be 2-4 folds larger than the target size without PVE, and that the Southampton method was remarkably accurate.

References
1.
Asenbaum S, Pirker W, Angelberger P, Bencsits G, Pruckmayer M, Brucke T . [123I]beta-CIT and SPECT in essential tremor and Parkinson's disease. J Neural Transm (Vienna). 1999; 105(10-12):1213-28. DOI: 10.1007/s007020050124. View

2.
Soret M, Koulibaly P, Darcourt J, Buvat I . Partial volume effect correction in SPECT for striatal uptake measurements in patients with neurodegenerative diseases: impact upon patient classification. Eur J Nucl Med Mol Imaging. 2006; 33(9):1062-72. DOI: 10.1007/s00259-005-0003-4. View

3.
Tossici-Bolt L, Dickson J, Sera T, Booij J, Asenbaun-Nan S, Bagnara M . [I]FP-CIT ENC-DAT normal database: the impact of the reconstruction and quantification methods. EJNMMI Phys. 2017; 4(1):8. PMC: 5272851. DOI: 10.1186/s40658-017-0175-6. View

4.
Booij J, Habraken J, Bergmans P, Tissingh G, Winogrodzka A, Wolters E . Imaging of dopamine transporters with iodine-123-FP-CIT SPECT in healthy controls and patients with Parkinson's disease. J Nucl Med. 1998; 39(11):1879-84. View

5.
Tossici-Bolt L, Hoffmann S, Kemp P, Mehta R, Fleming J . Quantification of [123I]FP-CIT SPECT brain images: an accurate technique for measurement of the specific binding ratio. Eur J Nucl Med Mol Imaging. 2006; 33(12):1491-9. DOI: 10.1007/s00259-006-0155-x. View