Errors in the Total Testing Process in the Clinical Chemistry Laboratory at the University of Gondar Hospital, Northwest Ethiopia

Overview
Specialty Health Services
Date 2018 Jul 10
PMID 29983521
Citations 9
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Laboratory services have been described as the major processes contributing to safe patient care in the modern healthcare sector. However, occurrences of errors in the overall testing processes impair the clinical decision-making process. Such errors are supposed to be high in resource-poor countries, like Ethiopia. The objective of this study was to assess errors in the total testing process in the Clinical Chemistry laboratory of the University of Gondar Hospital, Northwest Ethiopia.

Methods: A cross-sectional study was conducted at the University of Gondar Hospital from February to March 2016. All the required data were collected using established quality indicators. Data were analyzed using SPSS version 20. Frequencies and cross-tabulations were used to summarize descriptive statistics.

Results: A total of 3259 samples and corresponding laboratory request forms were received for analysis. The analysis of the overall distribution of errors revealed that 89.6% were pre-analytical errors, 2.6% were analytical, and 7.7% were post-analytical errors. Of the pre-analytical errors, incomplete request form filling was the most frequent error observed, followed by sample rejection rate (3.8%). Analytical errors related to internal and external quality control exceeding the target range, (14.4%) and (51.4%) respectively, were reported. Excessive turnaround time and unreported critical value cases were the major defects in the post-analytical phase of quality assurance.

Conclusion: The present finding showed relatively high frequency of errors, which alarms the importance of quality indicators to assess errors in the total testing process. The University of Gondar Hospital laboratory should improve the quality of healthcare services based on these findings using laboratory standards.

Citing Articles

Determining the rate and reasons for specimen rejection among specimens referred for HIV viral load determination through referral linkage to Debre Tabor Comprehensive Specialized Hospital, north-central, Ethiopia,2023: Retrospective study.

Berhan A, Almaw A, Damtie S, Solomon Y, Legese B, Getie B Heliyon. 2024; 10(11):e31736.

PMID: 38845923 PMC: 11153166. DOI: 10.1016/j.heliyon.2024.e31736.


Laboratory specimen rejection rate and associated factors among referred specimens at Debre Markos Referral Hospital, Ethiopia: prospective cross-sectional study.

Mesganaw B, Hassen F, Molla H, Misganaw K Pan Afr Med J. 2024; 47:112.

PMID: 38828425 PMC: 11143071. DOI: 10.11604/pamj.2024.47.112.33795.


Assessment of hematology laboratory performance in the total testing process using quality indicators and sigma metrics in the northwest of Ethiopia: A cross-sectional study.

Berta D, Grima M, Melku M, Adane T, Chane E, Teketelew B Health Sci Rep. 2024; 7(1):e1833.

PMID: 38264158 PMC: 10803892. DOI: 10.1002/hsr2.1833.


Magnitude of extra-analytical errors and associated factors in medical laboratories in thirteen Public Hospitals in Addis Ababa, Ethiopia.

Degfe C, Mohammed O SAGE Open Med. 2023; 11:20503121221148062.

PMID: 36643202 PMC: 9834788. DOI: 10.1177/20503121221148062.


Blood specimen rejection rate in clinical laboratory: A systematic review and meta-analysis.

Getawa S, Aynalem M, Melku M, Adane T Pract Lab Med. 2022; 33:e00303.

PMID: 36582816 PMC: 9792348. DOI: 10.1016/j.plabm.2022.e00303.


References
1.
Plebani M, Sciacovelli L, Marinova M, Marcuccitti J, Chiozza M . Quality indicators in laboratory medicine: a fundamental tool for quality and patient safety. Clin Biochem. 2012; 46(13-14):1170-4. DOI: 10.1016/j.clinbiochem.2012.11.028. View

2.
Goswami B, Singh B, Chawla R, Mallika V . Evaluation of errors in a clinical laboratory: a one-year experience. Clin Chem Lab Med. 2010; 48(1):63-6. DOI: 10.1515/CCLM.2010.006. View

3.
Stark A, Jones B, Chapman D, Well K, Krajenta R, Meier F . Clinical laboratory specimen rejection--association with the site of patient care and patients' characteristics: findings from a single health care organization. Arch Pathol Lab Med. 2007; 131(4):588-92. DOI: 10.5858/2007-131-588-CLSRWT. View

4.
Atay A, Demir L, Cuhadar S, Saglam G, Unal H, Aksun S . Clinical biochemistry laboratory rejection rates due to various types of preanalytical errors. Biochem Med (Zagreb). 2014; 24(3):376-82. PMC: 4210258. DOI: 10.11613/BM.2014.040. View

5.
Ricos C, Garcia-Victoria M, De la Fuente B . Quality indicators and specifications for the extra-analytical phases in clinical laboratory management. Clin Chem Lab Med. 2004; 42(6):578-82. DOI: 10.1515/CCLM.2004.100. View