» Articles » PMID: 29945929

Minimally Invasive Implantation of a Micropacemaker Into the Pericardial Space

Overview
Date 2018 Jun 28
PMID 29945929
Citations 6
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Permanent cardiac pacemakers require invasive procedures with complications often related to long pacemaker leads. We are developing a percutaneous pacemaker for implantation of an entire pacing system into the pericardial space.

Methods: Percutaneous micropacemaker implantations were performed in 6 pigs (27.4-34.1 kg) using subxyphoid access to the pericardial space. Modifications in the implantation methods and hardware were made after each experiment as the insertion method was optimized. In the first 5 animals, nonfunctional pacemaker devices were studied. In the final animal, a functional pacemaker was implanted.

Results: Successful placement of the entire nonfunctional pacing system into the pericardial space was demonstrated in 2 of the first 5 animals, and successful implantation and capture was achieved using a functional system in the last animal. A sheath was developed that allows retractable features to secure positioning within the pericardial space. In addition, a miniaturized camera with fiberoptic illumination allowed visualization of the implantation site before electrode insertion into myocardium. All animals studied during follow-up survived without symptoms after the initial postoperative period.

Conclusions: A novel micropacemaker system allows cardiac pacing without entering the vascular space or surgical exposure of the heart. This pericardial pacemaker system may be an option for a large number of patients currently requiring transvenous pacemakers but is particularly relevant for patients with restricted vascular access, young children, or those with congenital heart disease who require epicardial access.

Citing Articles

Precision imaging of cardiac function and scar size in acute and chronic porcine myocardial infarction using ultrahigh-field MRI.

Lohr D, Kollmann A, Bille M, Terekhov M, Elabyad I, Hock M Commun Med (Lond). 2024; 4(1):146.

PMID: 39026075 PMC: 11258271. DOI: 10.1038/s43856-024-00559-y.


MiRNA21 and IL-18 levels in left atrial blood in patients with atrial fibrillation undergoing cryoablation and their predictive value for recurrence of atrial fibrillation.

Liu Y, Luo D, Liu E, Liu T, Xu G, Liang X J Interv Card Electrophysiol. 2022; 64(1):111-120.

PMID: 35029768 DOI: 10.1007/s10840-022-01125-z.


Expanding the Reach of Pediatric Transcatheter Pacing.

J Innov Card Rhythm Manag. 2021; 12(4):4487-4492.

PMID: 33939789 PMC: 8081454. DOI: 10.19102/icrm.2021.120408.


A Multi-Dimensional Analysis of a Novel Approach for Wireless Stimulation.

Abiri P, Yousefi A, Abiri A, Gudapati V, Ding Y, Nguyen K IEEE Trans Biomed Eng. 2020; 67(12):3307-3316.

PMID: 32248088 PMC: 8262529. DOI: 10.1109/TBME.2020.2983443.


Chronic performance of subxiphoid minimally invasive pericardial Model 20066 pacemaker lead insertion in an infant animal model.

Clark B, Kumthekar R, Mass P, Opfermann J, Berul C J Interv Card Electrophysiol. 2019; 59(1):13-19.

PMID: 31612301 PMC: 7273738. DOI: 10.1007/s10840-019-00626-8.


References
1.
Horenstein M, Hakimi M, Walters 3rd H, Karpawich P . Chronic performance of steroid-eluting epicardial leads in a growing pediatric population: a 10-year comparison. Pacing Clin Electrophysiol. 2003; 26(7 Pt 1):1467-71. DOI: 10.1046/j.1460-9592.2003.t01-1-00212.x. View

2.
Loeb G, Zhou L, Zheng K, Nicholson A, Peck R, Krishnan A . Design and testing of a percutaneously implantable fetal pacemaker. Ann Biomed Eng. 2012; 41(1):17-27. PMC: 3524376. DOI: 10.1007/s10439-012-0631-3. View

3.
Clark B, Davis T, El-Sayed Ahmed M, McCarter R, Ishibashi N, Jordan C . Minimally invasive percutaneous pericardial ICD placement in an infant piglet model: Head-to-head comparison with an open surgical thoracotomy approach. Heart Rhythm. 2015; 13(5):1096-1104. DOI: 10.1016/j.hrthm.2015.12.015. View

4.
Janousek J, van Geldorp I, Krupickova S, Rosenthal E, Nugent K, Tomaske M . Permanent cardiac pacing in children: choosing the optimal pacing site: a multicenter study. Circulation. 2013; 127(5):613-23. DOI: 10.1161/CIRCULATIONAHA.112.115428. View

5.
Fortescue E, Berul C, Cecchin F, Walsh E, Triedman J, Alexander M . Comparison of modern steroid-eluting epicardial and thin transvenous pacemaker leads in pediatric and congenital heart disease patients. J Interv Card Electrophysiol. 2005; 14(1):27-36. DOI: 10.1007/s10840-005-3797-x. View