CT Pulmonary Angiography: Dose Reduction Via a Next Generation Iterative Reconstruction Algorithm

Overview
Journal Acta Radiol
Specialty Radiology
Date 2018 Jun 24
PMID 29933714
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Computed tomography pulmonary angiography (CTPA) is the standard imaging modality for detection or rule out of pulmonary embolism (PE); however, radiation exposure is a serious concern. With iterative reconstruction algorithms a distinct dose reduction could be achievable.

Purpose: To evaluate a next generation iterative reconstruction algorithm for detection or rule-out of PE in simulated low-dose CTPA.

Material And Methods: Low-dose CT datasets with 50%, 25%, and 12.5% of the original tube current were simulated based on CTPA examinations of 92 patients with suspected PE. All datasets were reconstructed with two reconstruction algorithms: standard filtered back-projection (FBP) and iterative model reconstruction (IMR). In total, 736 CTPA datasets were evaluated by three blinded radiologists regarding image quality, diagnostic confidence, and detectability of PE. Furthermore, contrast-to-noise ratio (CNR) was calculated.

Results: Images reconstructed with IMR showed better detectability of PE than images reconstructed with FBP, especially at lower dose levels. With IMR, sensitivity was over 95% for central and segmental PE down to a dose level of 25%. Significantly higher subjective image quality was shown at lower dose levels (25% and 12.5%) for IMR images whereas it was higher for FBP images at higher dose levels. FBP was rated as showing less artificial image appearance. CNR was significantly higher with IMR at all dose levels.

Conclusion: By using IMR, a dose reduction of up to 50% while maintaining satisfactory image quality seems feasible in standard clinical situations, resulting in a mean effective dose of 1.38 mSv for CTPA.

Citing Articles

Pregnancy-Related Cardiovascular Diseases: A Radiological Overview.

Lauriero F, Mazza G, Perazzolo A, Ottoni G, Cipriani A, Castro Pereira J J Cardiovasc Dev Dis. 2025; 12(2).

PMID: 39997477 PMC: 11856395. DOI: 10.3390/jcdd12020043.


Feasibility of Low-Dose and Low-Contrast Media Volume Approach in Computed Tomography Cardiovascular Imaging Reconstructed with Model-Based Algorithm.

Ippolito D, Porta M, Maino C, Riva L, Ragusi M, Giandola T Tomography. 2024; 10(2):286-298.

PMID: 38393291 PMC: 10891780. DOI: 10.3390/tomography10020023.


A closer look at the utilized radiation doses during computed tomography pulmonary angiography (CTPA) for COVID-19 patients.

Abuzaid M, Elshami W, Cavli B, Ozturk C, Almisned G, Tekin H Radiat Phys Chem Oxf Engl 1993. 2023; 211:111025.

PMID: 37250685 PMC: 10210819. DOI: 10.1016/j.radphyschem.2023.111025.


EFFICIENCY OF BREAST SHIELDING DEVICE IN OUT-OF-PLANE COMPUTERIZED TOMOGRAPHY IMAGING OF THE ABDOMEN - PRELIMINARY RESULTS.

Leder N, Popic J, Knezevic Z, Vidjak V Acta Clin Croat. 2023; 61(2):257-264.

PMID: 36818926 PMC: 9934029. DOI: 10.20471/acc.2022.61.02.12.


Split-bolus computed tomography urography (CTU) achieves more than half of radiation dose reduction in females and overweight patients than conventional single-bolus computed tomography urography.

Wang L, Wong Y, Hwang Y, Pang S, Chuang C, Chang Y Transl Oncol. 2021; 14(8):101151.

PMID: 34111709 PMC: 8193142. DOI: 10.1016/j.tranon.2021.101151.