» Articles » PMID: 29889983

A Model-based Tracking Method for Measuring 3D Dynamic Joint Motion Using an Alternating Biplane X-ray Imaging System

Overview
Journal Med Phys
Specialty Biophysics
Date 2018 Jun 12
PMID 29889983
Citations 9
Authors
Affiliations
Soon will be listed here.
Abstract

Purposes: To propose a new model-based tracking method for measuring three-dimensional (3D) dynamic joint kinematics using a clinical alternating biplane x-ray imaging system; and to quantify in vitro its errors in measuring ankle and knee motions at different motion speeds.

Methods: A new model-based tracking method based on motion component partition and interpolation (MCPI) was developed for measuring 3D dynamic joint kinematics based on a clinical alternating biplane x-ray imaging system. Two detectors of the biplane imaging system placed perpendicular to each other were operated to collect alternating fluoroscopic images of the targeted joint during tasks. The CT data of the joint were also acquired for the reconstruction of volumetric and surface models of each of the associated bones. The CT-based models of the bones were first registered to the alternating images using a model-to-single-plane fluoroscopic image registration method, and the resulting bone poses were then refined using a two-level optimization with motion component partition and model vertex trajectory interpolation. The MCPI method was evaluated in vitro for measurement errors for an ankle and a knee specimen moving at different speeds against a standard reference provided by a highly accurate motion capture system. The positional and rotational errors of the measured bone poses were quantified in terms of the bias, precision, and root-mean-squared errors (RMSE), as well as the mean target registration error (mTRE), a final mTRE less than 2.5 mm indicating a successful registration.

Results: The new method was found to have RMSE of bone pose measurements of less than 0.18 mm for translations and 0.72° for rotations for the ankle, and 0.33 mm and 0.74° for the knee with a high successful registration rate (>97%), and did not appear to be affected by joint motion speeds. Given the same alternating fluoroscopic images, the MCPI method outperformed the typical biplane analysis method assuming zero time offset between the two fluoroscopic views. The differences in performance between the methods were increased with increased joint motion speed. With the accurate bone pose data, the new method enabled talocrural, subtalar, and tibiofemoral kinematics measurements with submillimeter and subdegree accuracy, except for an RMSE of 1.04° for the internal/external rotation of the talocrural joint.

Conclusions: A new model-based tracking method based on MCPI has been developed for measuring dynamic joint motions using an alternating biplane x-ray imaging system widely available in medical centers. The MCPI method has been demonstrated in vitro to be highly accurate in determining the 3D kinematics of the bones of both the ankle joint complex and the knee. The current results suggest that the MCPI method would be an effective approach for measuring in vivo 3D kinematics of dynamic joint motion in a clinical setting equipped with an alternating biplane x-ray imaging system.

Citing Articles

Residual kinematic deviations of the shoulder during humeral elevation after conservative treatment for mid-shaft clavicle fractures.

Hung L, Lu H, Chen T, Wang T, Lu T Front Bioeng Biotechnol. 2024; 12:1413679.

PMID: 39183820 PMC: 11341403. DOI: 10.3389/fbioe.2024.1413679.


Integration of statistical shape modeling and alternating interpolation-based model tracking technique for measuring knee kinematics using clinical interleaved bi-plane fluoroscopy.

Lu H, Lin C, Shih K, Lu T, Kuo M, Li S PeerJ. 2023; 11:e15371.

PMID: 37334125 PMC: 10276557. DOI: 10.7717/peerj.15371.


Effects of Internal Fixation for Mid-Shaft Clavicle Fractures on Shoulder Kinematics During Humeral Elevations.

Hung L, Lu H, Chang C, Chen T, Wang T, Lu T Front Bioeng Biotechnol. 2021; 9:710787.

PMID: 34368104 PMC: 8339802. DOI: 10.3389/fbioe.2021.710787.


[Progress on in vivo ankle biomechanics based on dual fluoroscopic imaging technology].

Ye D, Sun X, Zhang C, Zhang S, Zhang X, Fu W Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2021; 38(3):602-608.

PMID: 34180207 PMC: 9927769. DOI: 10.7507/1001-5515.202006009.


Copers and Noncopers Use Different Landing Techniques to Limit Anterior Tibial Translation After Anterior Cruciate Ligament Reconstruction.

Keizer M, Otten E, Beijersbergen C, Brouwer R, Hijmans J Orthop J Sports Med. 2021; 9(4):2325967121998061.

PMID: 33948445 PMC: 8053773. DOI: 10.1177/2325967121998061.