Background: The rate of revision for some designs of total hip replacements due to idiopathic aseptic loosening has been reported as higher for women. However, whether this is environmental or inherently sex-related is not clear.

Objective: Can particle induced osteolysis be sex dependent? And if so, is this dependent on the type of implant debris (. metal polymer)? The objective of this study was to test for material dependent inflammatory osteolysis that may be linked to sex using CoCrMo and implant grade conventional polyethylene (UHMWPE), using an murine calvaria model.

Methods: Healthy 12 week old female and male C57BL/6J mice were treated with UHMWPE (1.0um ECD) or CoCrMo particles (0.9um ECD) or received sham surgery. Bone resorption was assessed by micro-computed tomography, histology and histomorphometry on day 12 post challenge.

Results: Female mice that received CoCrMo particles showed significantly more inflammatory osteolysis and bone destruction compared to the females who received UHMWPE implant debris. Moreover, females challenged with CoCrMo particles exhibited 120% more inflammatory bone loss compared to males (p<0.01) challenged with CoCrMo implant debris (but this was not the case for UHMWPE particles).

Conclusion: We demonstrated sex-specific differences in the amount of osteolysis resulting from CoCrMo particle challenge. This suggests osteo-immune responses to metal debris are preferentially higher in female compared to male mice, and supports the contention that there may be inherent sex related susceptibility to some types of implant debris.

Citing Articles

Hyaluronic acid-British anti-Lewisite as a safer chelation therapy for the treatment of arthroplasty-related metallosis.

Ude C, Schmidt S, Laurencin S, Shah S, Esdaille J, Kan H Proc Natl Acad Sci U S A. 2023; 120(45):e2309156120.

PMID: 37903261 PMC: 10636327. DOI: 10.1073/pnas.2309156120.


MAVRIC based T2 mapping assessment of infrapatellar fat pad scarring in patients with total knee arthroplasty.

Sacher S, Neri J, Gao M, Argentieri E, Potter H, Koch K J Orthop Res. 2022; 41(6):1299-1309.

PMID: 36262013 PMC: 10113607. DOI: 10.1002/jor.25472.


Sex-Specific Response to Combinations of Shear Stress and Substrate Stiffness by Endothelial Cells In Vitro.

James B, Allen J Adv Healthc Mater. 2021; 10(18):e2100735.

PMID: 34142471 PMC: 8458248. DOI: 10.1002/adhm.202100735.


Metal-induced delayed type hypersensitivity responses potentiate particle induced osteolysis in a sex and age dependent manner.

Samelko L, Caicedo M, McAllister K, Jacobs J, Hallab N PLoS One. 2021; 16(5):e0251885.

PMID: 34003868 PMC: 8130946. DOI: 10.1371/journal.pone.0251885.

References
1.
Caicedo M, Pennekamp P, McAllister K, Jacobs J, Hallab N . Soluble ions more than particulate cobalt-alloy implant debris induce monocyte costimulatory molecule expression and release of proinflammatory cytokines critical to metal-induced lymphocyte reactivity. J Biomed Mater Res A. 2009; 93(4):1312-21. DOI: 10.1002/jbm.a.32627. View

2.
Zhang K, Yang S, Yang S, Bai L, Li P, Liu D . Different influence of Ti, PMMA, UHMWPE, and Co-Cr particles on peripheral blood monocytes during periprosthetic inflammation. J Biomed Mater Res A. 2014; 103(1):358-64. DOI: 10.1002/jbm.a.35176. View

3.
Hallab N, McAllister K, Brady M, Jarman-Smith M . Macrophage reactivity to different polymers demonstrates particle size- and material-specific reactivity: PEEK-OPTIMA(®) particles versus UHMWPE particles in the submicron, micron, and 10 micron size ranges. J Biomed Mater Res B Appl Biomater. 2011; 100(2):480-92. DOI: 10.1002/jbm.b.31974. View

4.
Reddy A, Caicedo M, Samelko L, Jacobs J, Hallab N . Implant debris particle size affects serum protein adsorption which may contribute to particle size-based bioreactivity differences. J Long Term Eff Med Implants. 2014; 24(1):77-88. PMC: 4062925. DOI: 10.1615/jlongtermeffmedimplants.2014010118. View

5.
Chen Y, Hallab N, Liao Y, Narayan V, Schwarz E, Xie C . Antioxidant impregnated ultra-high molecular weight polyethylene wear debris particles display increased bone remodeling and a superior osteogenic:osteolytic profile vs. conventional UHMWPE particles in a murine calvaria model. J Orthop Res. 2015; 34(5):845-51. PMC: 4842171. DOI: 10.1002/jor.23080. View