Near-Continuous Glucose Monitoring Makes Glycemic Control Safer in ICU Patients
Overview
Emergency Medicine
Authors
Affiliations
Objectives: Tight glycemic control using intermittent blood glucose measurements is associated with a risk of hypoglycemia. Glucose concentrations can now be measured near continuously (every 5-15 min). We assessed the quality and safety of glycemic control guided by a near-continuous glucose monitoring system in ICU patients.
Design: Prospective, cluster-randomized, crossover study.
Setting: Thirty-five-bed medico-surgical department of intensive care with four separate ICUs.
Patients: Adult patients admitted to the department and expected to stay for at least 3 days were considered for inclusion if they had persistent hyperglycemia (blood glucose > 150 mg/dL) up to 6 hours after admission and/or were receiving insulin therapy.
Interventions: A peripheral venous catheter was inserted in all patients and connected to a continuous glucose monitoring sensor (GlucoClear; Edwards Lifesciences, Irvine, CA). The four ICUs were randomized in pairs in a crossover design to glycemic control using unblinded or blinded continuous glucose monitoring monitors. The insulin infusion rate was adjusted to keep blood glucose between 90 and 150 mg/dL using the blood glucose values displayed on the continuous glucose monitor (continuous glucose monitoring group-unblinded units) or according to intermittent blood glucose readings (intermittent glucose monitoring group-blinded units).
Measurements And Main Results: The quality and safety of glycemic control were assessed using the proportion of time in range, the frequency of blood glucose less than 70 mg/dL, and the time spent with blood glucose less than 70 mg/dL (TB70), using blood glucose values measured by the continuous glucose monitoring device. Seventy-seven patients were enrolled: 39 in the continuous glucose monitoring group and 38 in the intermittent glucose monitoring group. A total of 43,107 blood glucose values were recorded. The time in range was similar in the two groups. The incidence of hypoglycemia (8/39 [20.5%] vs 15/38 [39.5%]) and the TB70 (0.4% ± 0.9% vs 1.6% ± 3.4%; p < 0.05) was lower in the continuous glucose monitoring than in the intermittent glucose monitoring group.
Conclusions: Use of a continuous glucose monitoring-based strategy decreased the incidence and severity of hypoglycemia, thus improving the safety of glycemic control.
Accuracy of continuous glucose monitoring systems in intensive care unit patients: a scoping review.
Nielsen C, Grigonyte-Daraskeviciene M, Olsen M, Moller M, Norgaard K, Perner A Intensive Care Med. 2024; 50(12):2005-2018.
PMID: 39417874 DOI: 10.1007/s00134-024-07663-6.
Wang Y, Li S, Lu J, Feng K, Huang X, Hu F J Endocrinol Invest. 2024; 47(12):3091-3099.
PMID: 38762634 PMC: 11549136. DOI: 10.1007/s40618-024-02393-4.
Wang Y, Li S, Lu J, Feng K, Huang X, Hu F Diabetologia. 2024; 67(7):1295-1303.
PMID: 38568252 PMC: 11153265. DOI: 10.1007/s00125-024-06136-1.
Continuous Glucose Monitoring in the Intensive Care Unit.
Guerrero-Arroyo L, Faulds E, Perez-Guzman M, Davis G, Dungan K, Pasquel F J Diabetes Sci Technol. 2023; 17(3):667-678.
PMID: 37081830 PMC: 10210113. DOI: 10.1177/19322968231169522.
Martensson J, Cutuli S, Yanase F, Ancona P, Toh L, Osawa E Acta Anaesthesiol Scand. 2022; 67(1):86-93.
PMID: 36263915 PMC: 10092568. DOI: 10.1111/aas.14159.