Magnitude of Cone Beam CT Image Artifacts Related to Zirconium and Titanium Implants: Impact on Image Quality

Overview
Date 2018 Apr 19
PMID 29668300
Citations 33
Authors
Affiliations
Soon will be listed here.
Abstract

Objectives: To evaluate the magnitude of artifacts related to titanium and zirconium implants at different distances and angulations and their impact on cone beam CT(CBCT) image quality.

Methods: CBCT images were obtained before and after the insertion of titanium and zirconium implants in a mandible on different CBCT units: Picasso Trio, ProMax 3D and 3D Accuitomo 80. Artifact was assessed by measuring the standard deviation (SD) of gray values and contrast-to-noise ratio (CNR) of 11 regions of interest (ROIs) at different distances (1.5 cm, 2.5 cm and 3.5 cm) and angulations (65°, 90°, 115° and 140°) from implant region.

Results: For titanium images, SD values did not differ from those of images without implant in all ROIs; however, some effect occurred in Picasso images as higher values were observed in ROIs closer to the implant (p < 0.05). Zirconium images showed higher SD values than the others in some ROIs for Picasso and ProMax (p < 0.05). In ProMax, the difference was observed even in the farthest ROIs from the implant. CNR values were not influenced by the ROI in Picasso, but presented lower values in ROIs closer to the zirconium implant for ProMax and Accuitomo.

Conclusions: The quantity and magnitude of artifacts in CBCT are influenced by the type of implant and CBCT unit. Although they are more pronounced in regions closer to the implant and located at 90° in relation to the mandibular long axis, they can reach as far as 3.5 cm from the artifact-generator object.

Citing Articles

Evaluation of the effect of reducing metal artifacts in multi-detector CT imaging of zirconia and titanium implants.

Suito H, Yoshihara H, Maeda N, Kasai R, Inoue T, Amano M Oral Radiol. 2025; .

PMID: 40038165 DOI: 10.1007/s11282-025-00814-5.


Expression of beam hardening artifacts on horizontally stitched cone-beam computed tomography images.

Melo da Fonte J, Fontenele R, Freitas D Imaging Sci Dent. 2025; 54(4):327-335.

PMID: 39744557 PMC: 11685307. DOI: 10.5624/isd.20240061.


Dual-encoder architecture for metal artifact reduction for kV-cone-beam CT images in head and neck cancer radiotherapy.

Ki J, Lee J, Lee W, Kim J, Jin H, Jung S Sci Rep. 2024; 14(1):27907.

PMID: 39537735 PMC: 11561079. DOI: 10.1038/s41598-024-79305-2.


Combination of metal artifact reduction and sharpening filter application for horizontal root fracture diagnosis in teeth adjacent to a zirconia implant.

Ruiz D, de Oliveira Reis L, Fontenele R, Miranda-Viana M, Farias-Gomes A, Freitas D Imaging Sci Dent. 2024; 54(3):289-295.

PMID: 39371300 PMC: 11450410. DOI: 10.5624/isd.20240056.


The Impact of AI on Metal Artifacts in CBCT Oral Cavity Imaging.

Wajer R, Wajer A, Kazimierczak N, Wilamowska J, Serafin Z Diagnostics (Basel). 2024; 14(12).

PMID: 38928694 PMC: 11203150. DOI: 10.3390/diagnostics14121280.


References
1.
Schulze R, Berndt D, dHoedt B . On cone-beam computed tomography artifacts induced by titanium implants. Clin Oral Implants Res. 2009; 21(1):100-7. DOI: 10.1111/j.1600-0501.2009.01817.x. View

2.
Prell D, Kyriakou Y, Beister M, Kalender W . A novel forward projection-based metal artifact reduction method for flat-detector computed tomography. Phys Med Biol. 2009; 54(21):6575-91. DOI: 10.1088/0031-9155/54/21/009. View

3.
Oenning A, Jacobs R, Pauwels R, Stratis A, Hedesiu M, Salmon B . Cone-beam CT in paediatric dentistry: DIMITRA project position statement. Pediatr Radiol. 2017; 48(3):308-316. DOI: 10.1007/s00247-017-4012-9. View

4.
Vasconcelos T, Bechara B, McMahan C, Freitas D, Noujeim M . Evaluation of artifacts generated by zirconium implants in cone-beam computed tomography images. Oral Surg Oral Med Oral Pathol Oral Radiol. 2017; 123(2):265-272. DOI: 10.1016/j.oooo.2016.10.021. View

5.
Bechara B, McMahan C, Moore W, Noujeim M, Geha H . Contrast-to-noise ratio with different large volumes in a cone-beam computerized tomography machine: an in vitro study. Oral Surg Oral Med Oral Pathol Oral Radiol. 2012; 114(5):658-65. DOI: 10.1016/j.oooo.2012.08.436. View