» Articles » PMID: 29582811

Application of High Intensity Focused Ultrasound for Treatment of Open-angle Glaucoma in Indian Patients

Overview
Specialty Ophthalmology
Date 2018 Mar 28
PMID 29582811
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: The aim of this study is to assess the efficacy of ultrasound cycloplasty (UCP) in Indian patients with open angle glaucoma (OAG).

Methods: A prospective interventional study was designed to treat 73 eyes of 73 patients with OAG with the EyeOP1 device equipped with six miniaturized cylindrical piezoelectric transducers. Two treatment protocols of ultrasound delivery depending on exposure time (8 s and 10 s of shot per transducer) were used. Complete ophthalmic examination, ocular biometry and anterior segment optical coherence tomography were performed preoperatively and patients were followed up at day 1, day 7 and months 1, 2, 3, 6, and 12. The primary outcome measure was "successful" intraocular pressure (IOP) control defined as IOP reduction ≥20% from baseline and IOP value >5 mm Hg at the last follow-up visit. Secondary outcomes were the occurrence of complications and mean IOP during the follow-up period.

Results: In all patients, the mean IOP reduced from 23.5 ± 3.0 mmHg before treatment to 15.7 ± 5.4 mmHg at 12 months (P < 0.05). Successful IOP control after a single procedure was 78.3% (79% and 78% in the 8 s and 10 s groups, respectively) at 12 months. Overall, the mean IOP reduction achieved in responding patients was 41% (standard deviation = 12%). Notwithstanding minor side effects such as transient pain, anterior chamber reaction, and refractive error changes, no major intraoperative or postoperative complications (severe hypotony or phthisis) were observed during the follow-up.

Conclusion: Our short-term results reveal that UCP is a simple, safe, and noninvasive procedure which enables to significantly reduce the IOP in patients with OAG. The study results in Indian eyes corroborate findings in earlier studies on Caucasian eyes.

Citing Articles

Efficacy and safety of ultrasound cycloplasty for the treatment of glaucoma: a Meta-analysis.

Wu T, Yin X, Li J, Lu P Int J Ophthalmol. 2023; 16(8):1317-1325.

PMID: 37602344 PMC: 10398525. DOI: 10.18240/ijo.2023.08.19.


Commentary: High-intensity focused ultrasound - A panacea in the making?.

Behera G, Kaliaperumal S Indian J Ophthalmol. 2022; 70(12):4172-4173.

PMID: 36453308 PMC: 9940530. DOI: 10.4103/ijo.IJO_2037_22.


Efficacy and safety of ultrasound cycloplasty in Indian eyes with open-angle glaucoma.

Paul C, Divya J, Sengupta S, Kamal R, Paul A, Mitra I Indian J Ophthalmol. 2022; 70(12):4168-4171.

PMID: 36453307 PMC: 9940526. DOI: 10.4103/ijo.IJO_827_22.


Efficacy and safety of high-intensity focused ultrasound cyclo-plasty in glaucoma.

Chen D, Guo X, Luo S, Lu Y, Tang X BMC Ophthalmol. 2022; 22(1):401.

PMID: 36207670 PMC: 9542456. DOI: 10.1186/s12886-022-02622-5.


Phaco-UCP; combined phacoemulsification and ultrasound ciliary plasty versus phacoemulsification alone for management of coexisting cataract and open angle glaucoma: a randomized clinical trial.

Torky M, Alzafiri Y, Abdelhameed A, Awad E BMC Ophthalmol. 2021; 21(1):53.

PMID: 33478426 PMC: 7819220. DOI: 10.1186/s12886-021-01818-5.


References
1.
Fankhauser F, Kwasniewska S, van der Zypen E . Cyclodestructive procedures. I. Clinical and morphological aspects: a review. Ophthalmologica. 2004; 218(2):77-95. DOI: 10.1159/000076142. View

2.
Ramli N, Htoon H, Ho C, Aung T, Perera S . Risk factors for hypotony after transscleral diode cyclophotocoagulation. J Glaucoma. 2011; 21(3):169-73. DOI: 10.1097/IJG.0b013e318207091a. View

3.
Tham Y, Li X, Wong T, Quigley H, Aung T, Cheng C . Global prevalence of glaucoma and projections of glaucoma burden through 2040: a systematic review and meta-analysis. Ophthalmology. 2014; 121(11):2081-90. DOI: 10.1016/j.ophtha.2014.05.013. View

4.
Giannaccare G, Vagge A, Gizzi C, Bagnis A, Sebastiani S, Del Noce C . High-intensity focused ultrasound treatment in patients with refractory glaucoma. Graefes Arch Clin Exp Ophthalmol. 2016; 255(3):599-605. DOI: 10.1007/s00417-016-3563-z. View

5.
Aptel F, Dupuy C, Rouland J . Treatment of refractory open-angle glaucoma using ultrasonic circular cyclocoagulation: a prospective case series. Curr Med Res Opin. 2014; 30(8):1599-605. DOI: 10.1185/03007995.2014.910509. View